Boston Scientific Cork recalls
41 recent recalls on record for Boston Scientific Cork in connected official government data as of November 25, 2026 at 12:00 AM UTC.
Models with recalls
All Boston Scientific Cork recalls
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0263-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0262-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0261-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0260-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0259-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0258-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0257-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0256-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0255-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0254-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0253-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0252-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0251-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0250-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0249-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0248-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0247-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0246-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0245-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0244-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0243-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0242-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0241-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0240-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0239-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0238-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0237-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0236-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0235-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0234-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0233-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0232-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0231-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0230-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0229-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0228-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0227-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0226-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0225-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0224-2027 ·
Official record ↗
FDA Class II · Medical device · October 28, 2026
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridge…
Official ID Z-0223-2027 ·
Official record ↗
Compiled from official public-domain government data. Last checked November 25, 2026 at 12:00 AM UTC.