Fresh Recalls

Diagnostica Stago recalls

Recall found. Diagnostica Stago is named in FDA Class II recall Z-3047-2026 covering STA Staclot Protein S. Hazard: A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified Identifying details: UDI : (01)3607450007466(11)250331(17)270331(10)273074(241)00746, Exp 3/31/2027. Status: Ongoing.

Models with recalls

STA Staclot Protein S (1)

All Diagnostica Stago recalls

FDA Class II · Medical device · September 9, 2026

STA Staclot Protein S recall — Diagnostica Stago, Inc.

A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified

Official ID Z-3047-2026 · Official record ↗

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from official public-domain government data. Last checked October 22, 2026 at 6:00 PM UTC.