Fresh Recalls

Emano Metrics recalls

Recall found. Emano Metrics is named in FDA Class II recall Z-2777-2026 covering Emano Flow with software version v2. Hazard: Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams. Identifying details: Software Version - 2.0.0 GS1 GTIN code: 199874903484 UDI code: GUDID registration is in progress. Status: Ongoing.

Models with recalls

Emano Flow with software version v2 (1)

All Emano Metrics recalls

FDA Class II · Medical device · July 29, 2026

Emano Flow with software version v2 recall — Emano Metrics, Inc.

Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validat…

Official ID Z-2777-2026 · Official record ↗

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from official public-domain government data. Last checked September 7, 2026 at 6:00 AM UTC.