Erbe USA recalls
Recall found. Erbe USA is named in FDA Class II recall Z-2938-2026 covering ERBECRYO 2 Cryosurgical Unit Model/Catalog Number.
Hazard: A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture. Identifying details: UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3. Status: Ongoing.
Models with recalls
All Erbe USA recalls
FDA Class II · Medical device · August 19, 2026
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the r…
Official ID Z-2938-2026 ·
Official record ↗