Fresh Recalls

Erbe USA recalls

Recall found. Erbe USA is named in FDA Class II recall Z-2938-2026 covering ERBECRYO 2 Cryosurgical Unit Model/Catalog Number. Hazard: A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture. Identifying details: UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3. Status: Ongoing.

Models with recalls

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number (1)

All Erbe USA recalls

FDA Class II · Medical device · August 19, 2026

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number recall — Erbe USA Inc

A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the r…

Official ID Z-2938-2026 · Official record ↗

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from official public-domain government data. Last checked September 7, 2026 at 6:00 AM UTC.