GE Medical Systems Information Technologies recalls

14 recent recalls on record for GE Medical Systems Information Technologies in connected official government data as of October 22, 2026 at 6:00 PM UTC.

Models with recalls

CASE v7 (1)Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04 (1)FRU (Field replaceable unit) (1)GE HealthCare APEXPRO v5 (1)GE HealthCare ApexPro v6 (1)GE HealthCare CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA (2)GE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER (1)GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA (1)GE Healthcare MAC VU360 (1)GE Healthcare Portrait Core Services SW V 1 (1)PORTRAIT COMPUTE AND CORE SERVICES V1 (1)PORTRAIT CORE SERVICES SOFTWARE V1 (2)

All GE Medical Systems Information Technologies recalls

FDA Class II · Medical device · September 23, 2026

Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04 recall — GE Medical Systems Information Technologies Inc

Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

Official ID Z-3143-2026 · Official record ↗

FDA Class II · Medical device · September 2, 2026

FRU (Field replaceable unit) recall — GE Medical Systems Information Technologies Inc

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and…

Official ID Z-3039-2026 · Official record ↗

FDA Class II · Medical device · September 2, 2026

PORTRAIT CORE SERVICES SOFTWARE V1 recall — GE Medical Systems Information Technologies Inc

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and…

Official ID Z-3038-2026 · Official record ↗

FDA Class II · Medical device · September 2, 2026

PORTRAIT COMPUTE AND CORE SERVICES V1 recall — GE Medical Systems Information Technologies Inc

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and…

Official ID Z-3037-2026 · Official record ↗

FDA Class II · Medical device · September 2, 2026

PORTRAIT CORE SERVICES SOFTWARE V1 recall — GE Medical Systems Information Technologies Inc

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and…

Official ID Z-3036-2026 · Official record ↗

FDA Class II · Medical device · August 12, 2026

GE Healthcare Portrait Core Services SW V 1 recall — GE Medical Systems Information Technologies Inc

A software vulnerability may allow unauthorized persons to gain privileged access to the device, which could result in loss of patient monitoring, missed patient alarms, and/or acc…

Official ID Z-2878-2026 · Official record ↗

FDA Class II · Medical device · August 12, 2026

GE Healthcare MAC VU360 recall — GE Medical Systems Information Technologies Inc

Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of i…

Official ID Z-2876-2026 · Official record ↗

FDA Class II · Medical device · June 10, 2026

CASE v7 recall — GE Medical Systems Information Technologies Inc

Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charr…

Official ID Z-2279-2026 · Official record ↗

FDA Class II · Medical device · June 3, 2026

GE HealthCare ApexPro v6 recall — GE Medical Systems Information Technologies Inc

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result…

Official ID Z-2247-2026 · Official record ↗

FDA Class II · Medical device · June 3, 2026

GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA recall — GE Medical Systems Information Technologies Inc

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result…

Official ID Z-2246-2026 · Official record ↗

FDA Class II · Medical device · June 3, 2026

GE HealthCare CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA recall — GE Medical Systems Information Technologies Inc

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result…

Official ID Z-2245-2026 · Official record ↗

FDA Class II · Medical device · June 3, 2026

GE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER recall — GE Medical Systems Information Technologies Inc

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result…

Official ID Z-2244-2026 · Official record ↗

FDA Class II · Medical device · June 3, 2026

GE HealthCare APEXPRO v5 recall — GE Medical Systems Information Technologies Inc

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result…

Official ID Z-2243-2026 · Official record ↗

FDA Class II · Medical device · June 3, 2026

GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL recall — GE Medical Systems Information Technologies Inc

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result…

Official ID Z-2242-2026 · Official record ↗

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from official public-domain government data. Last checked October 22, 2026 at 6:00 PM UTC.