Medacta Usa recalls

Recall found. Medacta Usa is named in FDA Class II recall Z-2618-2026 covering Short Monobloc Stem Broaches. Hazard: Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus. Identifying details: REF/UDI/Lot: 04.01.10.0970/07630542768910/2585086, 04.01.10.0971/07630542768927/2585087, 04.01.10.0972/07630542768934/2585088, 04.01.10.0973/07630542768941/2585089, 04.01.10.0974/07630542768958/2585090, 04.01.10.0975/07630542768965/2585091, 04.01.10.0976/07630542768972/2585092, 04.01.10.0977/07630542768989/2585093, 04.01.10.0978/07630542768996/2585094, 04.01.10.0979/07630542769009/2585095. Status: Ongoing.

Models with recalls

Short Monobloc Stem Broaches (1)

All Medacta Usa recalls

FDA Class II · Medical device · July 8, 2026

Short Monobloc Stem Broaches recall — Medacta Usa Inc

Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are posit…

Official ID Z-2618-2026 · Official record ↗

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from official public-domain government data. Last checked October 22, 2026 at 6:00 PM UTC.