Recall found. Radiometer Medical ApS is named in FDA Class II recall Z-0190-2027 covering The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists. Hazard: In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings. Identifying details: UDI-DI: 05700693930923, 05700693930909 Software versions 3.6 and lower. ABL90 FLEX PLUS Instructions for Use versions prior to: 996-223, Version: 202606O. ABL90 FLEX Instructions for Use versions prior to: 996-189, Version: 202606L. Status: Ongoing.

Models with recalls

The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists (1)

All Radiometer Medical ApS recalls

FDA Class II · Medical device · October 21, 2026

The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists recall — Radiometer Medical ApS

In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer su…

Official ID Z-0190-2027 · Official record ↗