Recall found. ResMed is named in FDA Class I recall Z-2735-2026 covering Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous. Hazard: Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy. Identifying details: Catalog Number: Astral 100 UDI-DI code: 00619498003259 Serial Numbers: Astral Ventilators-Serial number < 22241890149 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070 Catalog Number: Astral 150 UDI-DI code: 00619498270033 Serial Numbers: Astral Ventilator-Serial numbers < 22241440282 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070. Status: Ongoing.

Models with recalls

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous (1)

All ResMed recalls

FDA Class I · Medical device · August 5, 2026

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous recall — ResMed Ltd.

Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.

Official ID Z-2735-2026 · Official record ↗