Recall found. Technological Medical Advancements is named in FDA Class II recall Z-2162-2026 covering Diowave Laser System.
Hazard: Software update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associated with minor arthritis, and to temporarily increase local blood circulation, operating at a maximum average power output of 70W, which is beyond the FDA cleared 60W. Identifying details: UDI-DI: G180DIOWAVESTEALTHMAX0. All software versions prior to, but not including V1.81. Status: Ongoing.
Models with recalls
All Technological Medical Advancements recalls
FDA Class II · Medical device · May 20, 2026
Software update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associated with minor arthri…
Official ID Z-2162-2026 ·
Official record ↗