Fresh Recalls

Terex Utilities recalls

Recall found. Terex Utilities is named in FDA Class II recall Z-2745-2026 covering Power Over Fiber Base Module. Hazard: Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning statement Identifying details: Power Over Fiber Base Module Identification P/N 633232 P/N 622365 P/N 633383 P/N 633910 P/N 635731. Status: Ongoing.

Models with recalls

Power Over Fiber Base Module (1)

All Terex Utilities recalls

FDA Class II · Medical device · August 5, 2026

Power Over Fiber Base Module recall — Terex Utilities

Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for…

Official ID Z-2745-2026 · Official record ↗

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from official public-domain government data. Last checked September 7, 2026 at 6:00 AM UTC.