The Binding Site Group recalls
Recall found. The Binding Site Group is named in FDA Class II recall Z-2362-2026 covering Brand Name.
Hazard: It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration. Identifying details: Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20.. Status: Ongoing.
Models with recalls
All The Binding Site Group recalls
FDA Class II · Medical device · June 17, 2026
It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clini…
Official ID Z-2362-2026 ·
Official record ↗
Compiled from official public-domain government data. Last checked October 22, 2026 at 6:00 PM UTC.