Fresh Recalls

FDAClass II · Medical device · Recall Z-0079-2027 · Recalled October 14, 2026

Presource Packs containing Medtronic Microstream Components recall — Cardinal Health 200, LLC

Agency report date: October 14, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Presource Packs containing Medtronic Microstream Components
Company / brand
Cardinal Health 200
Lot / serial / date codes
AN23366084 UDI-DI 10197106950074 Kit lot 40711 AN23366085 UDI-DI 10198956177987 Kit lot 46520 AN2336608H UDI-DI 10197106159828 Kit lots 20866 23139 24264 25795 27111 27899 29023 29488 30527 31037 31698 32777 33326 AN2336608J UDI-DI 10197106704035 Kit lots 33461 33624 35231 35535 35689 36554 36600 36825 38327 38891 40511 41864 43975 44933 45972 47547 47897 48088 48528 49440 OMSVM4166A UDI-DI 10198956162259 Kit Lots 47765 50521 53221 53929 54565 54644 AN2336608G UDI-DI 10197106052105 Kit lots 18464 18569 19224
Distribution
Pending
Agency product description and boundaries

Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366085 AN2336608H AN2336608J Intubating Bag Kit Catalog Number OMSVM4166A Intubation UH Kit Catalog Number AN2336608G

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Presource Packs containing Medtronic Microstream Components recall status

3. Open the official notice or contact

Official recall ID
Z-0079-2027
Brand
All Cardinal Health 200 recalls
Classification
Class II
Units affected
210 total kits
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Waukegan, IL, United States
Firm address
3651 Birchwood Dr, N/A, 60085-8337

Identifying details (verbatim from the notice)

Product codes / lots
AN23366084 UDI-DI 10197106950074 Kit lot 40711 AN23366085 UDI-DI 10198956177987 Kit lot 46520 AN2336608H UDI-DI 10197106159828 Kit lots 20866 23139 24264 25795 27111 27899 29023 29488 30527 31037 31698 32777 33326 AN2336608J UDI-DI 10197106704035 Kit lots 33461 33624 35231 35535 35689 36554 36600 36825 38327 38891 40511 41864 43975 44933 45972 47547 47897 48088 48528 49440 OMSVM4166A UDI-DI 10198956162259 Kit Lots 47765 50521 53221 53929 54565 54644 AN2336608G UDI-DI 10197106052105 Kit lots 18464 18569 19224
Distribution
Pending
Firm notified the public via
Letter
FDA event ID
99801
Recall initiated
August 19, 2026
FDA classified
October 2, 2026
FDA report date
October 14, 2026

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 12:00 PM UTC. Record retrieved October 22, 2026 at 12:00 PM UTC.