Fresh Recalls

FDAClass II · Medical device · Recall Z-0107-2027 · Recalled October 14, 2026

SEDIPLAST recall — POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC

Agency report date: October 14, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
SEDIPLAST
Company / brand
POLYMEDCO
Lot / serial / date codes
Model/Catalog No. S-100 and S-1000; UDI-DI: N/A; Lot Number E1150A; Expiration Date 06/30/27
Distribution
Worldwide - US Nationwide distribution in the states and territory of AK, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and the country of United Kingdom.
Agency product description and boundaries

SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: SEDIPLAST recall status

3. Open the official notice or contact

Official recall ID
Z-0107-2027
Brand
All POLYMEDCO recalls
Classification
Class II
Units affected
448,400
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Cortlandt Manor, NY, United States
Firm address
510 Furnace Dock Rd, N/A, 10567-6220

Identifying details (verbatim from the notice)

Product codes / lots
Model/Catalog No. S-100 and S-1000; UDI-DI: N/A; Lot Number E1150A; Expiration Date 06/30/27
Distribution
Worldwide - US Nationwide distribution in the states and territory of AK, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and the country of United Kingdom.
Firm notified the public via
E-Mail
FDA event ID
99810
Recall initiated
August 25, 2026
FDA classified
October 7, 2026
FDA report date
October 14, 2026

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 12:00 PM UTC. Record retrieved October 22, 2026 at 12:00 PM UTC.