Fresh Recalls

FDAClass I · Medical device · Recall Z-0128-2027 · Recalled October 28, 2026

VCSN+Pro Model Number recall — Ventec Life Systems, Inc.

Agency report date: October 28, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
VCSN+Pro Model Number
Company / brand
Ventec Life Systems
Lot / serial / date codes
All Serial Numbers: UDI: 00855573007860
Distribution
Distribution US nationwide including Puerto Rico.
Agency product description and boundaries

VCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class I: most serious — reasonable probability of serious health consequences or death.

Affected product: VCSN+Pro Model Number recall status

3. Open the official notice or contact

Official recall ID
Z-0128-2027
Brand
All Ventec Life Systems recalls
Classification
Class I
Units affected
4,368 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Bothell, WA, United States
Firm address
22002 26th Ave Se, 98021-4903

Identifying details (verbatim from the notice)

Product codes / lots
All Serial Numbers: UDI: 00855573007860
Distribution
Distribution US nationwide including Puerto Rico.
Firm notified the public via
Letter
FDA event ID
99835
Recall initiated
September 17, 2026
FDA classified
October 20, 2026
FDA report date
October 28, 2026

More Ventec Life Systems recalls

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked November 21, 2026 at 12:00 AM UTC. Record retrieved November 21, 2026 at 12:00 AM UTC.

Report a data error on this page