Fresh Recalls

⚠ Data freshness notice: the most recent refresh of NHTSA was incomplete, so a very recent recall may be missing. The stored records may also have changed at the agency since the last successful check; for anything time-critical, confirm with the official sources linked below.

FDAClass II · Medical device · Recall Z-0166-2027 · Recalled October 21, 2026

Triathlon Hinge Femoral Prep Tray recall — Howmedica Osteonics Corp.

Agency report date: October 21, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Triathlon Hinge Femoral Prep Tray
Company / brand
Howmedica Osteonics
Lot / serial / date codes
Catalog Number: 6543-6-514. GTIN: 07613327562859. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Distribution
US Nationwide distribution. International distribution to Canada, Australia, New Zealand.
Agency product description and boundaries

Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Triathlon Hinge Femoral Prep Tray recall status

3. Open the official notice or contact

Official recall ID
Z-0166-2027
Brand
All Howmedica Osteonics recalls
Classification
Class II
Units affected
340 trays
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Mahwah, NJ, United States
Firm address
325 Corporate Dr, 07430-2006

Identifying details (verbatim from the notice)

Product codes / lots
Catalog Number: 6543-6-514. GTIN: 07613327562859. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Distribution
US Nationwide distribution. International distribution to Canada, Australia, New Zealand.
Firm notified the public via
Letter
FDA event ID
99883
Recall initiated
September 16, 2026
FDA classified
October 15, 2026
FDA report date
October 21, 2026

More Howmedica Osteonics recalls

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked November 25, 2026 at 6:00 AM UTC. Record retrieved November 25, 2026 at 6:00 AM UTC.

Report a data error on this page