FDAClass II · Medical device · Recall Z-0197-2027 · Recalled October 28, 2026

Medtronic Biopsy Needle Kit recall — Medtronic Navigation, Inc.

Agency report date: October 28, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Medtronic Biopsy Needle Kit
Company / brand
Medtronic Navigation
Lot / serial / date codes
UDI: 00643169702943/ Lots: 066500526 066501226 066503326 066504726 066506326 066509826; UDI:00643169030121/Lots: 066500526 066506326
Distribution
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: Canary Islands, Iceland, Northern Ireland, Martinique, China , Taiwan, Hong Kong,
Agency product description and boundaries

Medtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, Rxonly

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Medtronic Biopsy Needle Kit recall status

3. Open the official notice or contact

Official recall ID
Z-0197-2027
Brand
All Medtronic Navigation recalls
Classification
Class II
Units affected
3746 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Lafayette, CO, United States
Firm address
200 Medtronic Dr, 80026-3735

Identifying details (verbatim from the notice)

Product codes / lots
UDI: 00643169702943/ Lots: 066500526 066501226 066503326 066504726 066506326 066509826; UDI:00643169030121/Lots: 066500526 066506326
Distribution
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: Canary Islands, Iceland, Northern Ireland, Martinique, China , Taiwan, Hong Kong,
Firm notified the public via
Letter
FDA event ID
99895
Recall initiated
August 31, 2026
FDA classified
October 16, 2026
FDA report date
October 28, 2026

More Medtronic Navigation recalls

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked November 25, 2026 at 6:00 AM UTC. Record retrieved November 25, 2026 at 6:00 AM UTC.

Report a data error on this page