FDAClass II · Medical device · Recall Z-0264-2027 · Recalled October 28, 2026
Generic Name recall — restor3d Inc.
Agency report date: October 28, 2026.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Agency product description and boundaries
Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 Software Version: None Product Description: The restor3d Metatarsophalangeal (MTP) Implant is a sterile, single-use implant grade metal device, available in varied articulating surface coverages. The restor3d MTP implant is designed to replace the damaged articulating surface of the first metatarsal distal head. The device is intended for the treatment of patients with degenerative and post-traumatic arthritis in the first metatarsal joint in the presence of good bone stock and integrity of the first metatarsal head. This implant can also be used in the following clinical conditions: hallux valgus or hallux limitus, hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The implant is a single use product intended to be press fit, with optional use of bone cement. The restor3d MTP Implant is comprised of a single laser powder bed fusion (LPBF) printed cobalt-chromium alloy (Co28Cr6Mo) part. The external surface of the implant that is intended to be in contact with the phalanx is mirror polished. Component: Not a component
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
2. Agency hazard and consumer action
FDA reports: Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.
No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.
Class II: may cause temporary or medically reversible health consequences.
Affected product: Generic Name recall status
3. Open the official notice or contact
View the official FDA enforcement record (openFDA) ↗
Affected products
Companies & origin
Identifying details (verbatim from the notice)
Related recalls
More restor3d recalls
- Veritas Reverse Total Shoulder Replacement recall — restor3d Inc.FDA · November 11, 2026
- Kinos Axiom Total Ankle System with Axiom PSR recall — restor3d Inc.FDA · September 23, 2026
- Brand Name recall — restor3d Inc.FDA · August 26, 2026
Other medical device recalls from around the same time
- Impella Optical Controller recall — Abiomed, Inc.FDA · November 4, 2026
- Automated Impella Controller recall — Abiomed, Inc.FDA · November 4, 2026
- Halyard ANTERIOR HIP PACK recall — AVID Medical, Inc.FDA · October 28, 2026
- Halyard ANTERIOR HIP PACK recall — AVID Medical, Inc.FDA · October 28, 2026
- MEDLINE medical convenience kits labeled as recall — Medline Industries, LPFDA · October 28, 2026
- CPSC — consumer products ↗ (aggregated here)
- NHTSA — vehicles, car seats, tires ↗ (aggregated here; use NHTSA's VIN lookup to check a specific vehicle)
- FDA — food, drugs, medical devices ↗ (aggregated here)
- USDA FSIS — meat, poultry & processed egg products ↗ (not aggregated here)
- Recalls.gov — the U.S. government’s all-agency portal ↗
Found something wrong or missing? Report a data error.
Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked November 25, 2026 at 6:00 AM UTC. Record retrieved November 25, 2026 at 6:00 AM UTC.