Fresh Recalls

FDAClass II · Medical device · Recall Z-0311-2027 · Recalled November 4, 2026

Nalu TD2 Therapy Disc with firmware recall — Boston Scientific Neuromodulation Corporation

Agency report date: November 4, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Nalu TD2 Therapy Disc with firmware
Company / brand
Boston Scientific Neuromodulation
Lot / serial / date codes
Firmware versions v2.0.71 and v2.0.77 REF/UDI-DI/Serial Numbers: 34014-001/00812537037134/ 100855, 100877, 100883, 100884, 100886, 100888, 100898, 100903, 100908, 100911, 100916, 100932, 101103, 100902; 74011/00812537037127, 00812537036977/101439, 100913, 100949, 101159, 101162, 101163, 101164, 101173, 101175, 101177, 101186, 101187, 101188, 101189, 101192, 101200, 101205, 101208, 101213, 101214, 101218, 101219, 101221, 101222, 101240, 101245, 101247, 101249, 101256, 101257, 101259, 101260, 101261, 101264, 101267, 101268, 101270, 101272, 101273, 101277, 101278, 101280, 101283, 101284, 101290, 101297, 101298, 101301, 101305, 101307, 101310, 101311, 101312, 101314, 101315, 101326, 101330, 101331, 101334, 101338, 101344, 101345, 101346, 101356, 101357, 101358, 101359, 101363, 101364, 101367, 101375, 101376, 101377, 101378, 101381, 101383, 101387, 101388, 101392, 101400, 101405, 101408, 101410, 101414, 101415, 101417, 101418, 101419, 101429, 101431, 101432, 101436, 101440, 101441, 101443, 101447, 101448, 101450, 101451, 101466, 101490, 101493, 101504, 101510, 101511, 101514, 101520, 101527, 101528, 101534, 101535, 101553, 101557, 101567, 101571, 101576, 101599, 101617, 101623, 101656
Distribution
US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.
Agency product description and boundaries

Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the Nalu Neurostimulation System

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Nalu TD2 Therapy Disc with firmware recall status

3. Open the official notice or contact

Official recall ID
Z-0311-2027
Brand
All Boston Scientific Neuromodulation recalls
Classification
Class II
Units affected
134
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Valencia, CA, United States
Firm address
25155 Rye Canyon Loop, 91355-5004

Identifying details (verbatim from the notice)

Product codes / lots
Firmware versions v2.0.71 and v2.0.77 REF/UDI-DI/Serial Numbers: 34014-001/00812537037134/ 100855, 100877, 100883, 100884, 100886, 100888, 100898, 100903, 100908, 100911, 100916, 100932, 101103, 100902; 74011/00812537037127, 00812537036977/101439, 100913, 100949, 101159, 101162, 101163, 101164, 101173, 101175, 101177, 101186, 101187, 101188, 101189, 101192, 101200, 101205, 101208, 101213, 101214, 101218, 101219, 101221, 101222, 101240, 101245, 101247, 101249, 101256, 101257, 101259, 101260, 101261, 101264, 101267, 101268, 101270, 101272, 101273, 101277, 101278, 101280, 101283, 101284, 101290, 101297, 101298, 101301, 101305, 101307, 101310, 101311, 101312, 101314, 101315, 101326, 101330, 101331, 101334, 101338, 101344, 101345, 101346, 101356, 101357, 101358, 101359, 101363, 101364, 101367, 101375, 101376, 101377, 101378, 101381, 101383, 101387, 101388, 101392, 101400, 101405, 101408, 101410, 101414, 101415, 101417, 101418, 101419, 101429, 101431, 101432, 101436, 101440, 101441, 101443, 101447, 101448, 101450, 101451, 101466, 101490, 101493, 101504, 101510, 101511, 101514, 101520, 101527, 101528, 101534, 101535, 101553, 101557, 101567, 101571, 101576, 101599, 101617, 101623, 101656
Distribution
US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.
FDA event ID
99785
Recall initiated
September 30, 2026
FDA classified
October 29, 2026
FDA report date
November 4, 2026

More Boston Scientific Neuromodulation recalls

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked November 21, 2026 at 12:00 AM UTC. Record retrieved November 21, 2026 at 12:00 AM UTC.

Report a data error on this page