FreshRecalls

FDAClass II · Medical device · Recalled May 6, 2026

Philips SmartPath to dStream for XR and 3 recall — Philips North America

Hazard: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Class II: may cause temporary or medically reversible health consequences.

Official recall ID
Z-1967-2026
Brand
Philips North America
Classification
Class II
Units affected
14 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Description

Philips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.

Affected products

Companies & origin

Recalling firm location
Cambridge, MA, United States
Firm address
222 Jacobs St, 02141-2296

Identifying details (verbatim from the notice)

Product codes / lots
1. Model Number (REF): 781270. UDI: (01)00884838095083(21)38109, (01)00884838095083(21)17386, (01)00884838095083(21)38356, (01)00884838095083(21)34211, (01)00884838095083(21)38113, (01)00884838095083(21)17356, (01)00884838095083(21)38111, (01)00884838095083(21)74011, (01)00884838095083(21)24058, (01)00884838095083(21)24066, (01)00884838095083(21)34223. Serial Numbers: 38109, 17386, 38356, 34211, 38113, 17356, 38111, 74011, 24058, 24066, 34223. 2. Model Number (REF): 782113. UDI: (01)00884838098909(21)38066, (01)00884838098909(21)38117, (01)00884838098909(21)17384. Serial Numbers: 38066, 38117, 17384.
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.
Firm notified the public via
Letter
FDA event ID
98779
Recall initiated
April 14, 2026
FDA classified
April 27, 2026
FDA report date
May 6, 2026

Source: U.S. Food & Drug Administration. This page mirrors the official public-domain record verbatim; the linked government source is authoritative. Source last checked July 24, 2026 at 3:00 AM UTC.