Fresh Recalls

FDAClass II · Medical device · Recall Z-2049-2026 · Recalled May 13, 2026

LUX-Dx II Plus recall — Boston Scientific Corporation

Agency report date: May 13, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
LUX-Dx II Plus
Company / brand
Boston Scientific
Lot / serial / date codes
LUX-Dx II Plus UDI-DI: 00802526620706 (US), 00802526623707 (EU/AU); SERVER SW LATITUDE DRAGON EU UDI-DI: 00802526613821; SERVER SW LATITUDE DRAGON AU UDI-DI: 00802526613814; Serial Number: 100085 100676 100909 100944 101811 102341 102405 102864 102956 102978 103018 105283 105697 106372 106429 106690 107386 107460 108132 108429 108433 108641 108697 108805 108861 108901 109663 109699 109836 110700 110737 110858 111048 111119 111269 111290 111362 111780 111987 112265 112549 113400 113495 113720 114072 114393 115310 115411 115835 115862 115890 115981 116066 116090 116197 117180 117533 117560 117820 117924 117960 119462 120931 121080 121203 121267 121483 122855 123305 123412 124378 124458 124557 124791 124987 125008 126249 127182 127441 127758 127821 128106 128506 129528 129763 129782 130474 131469 132487 132549 133195 133786 133989 134600 135362 136038 136746 136884 137047 137134 137319 137449 138039 138059 138334 138759 139080 139165 140319 141181 142273 143441 143887 144590 144777 145284 145336 145684 146709 147185 147618 148972 149028 153330 153628 155199 155735 155898 156145 156895 157872 158926 159162 160236 160471 160615 161307 161762 162801 162821 163556 163802 167866 169673 169701 171284 172766 175538
Distribution
Worldwide distribution - US Nationwide and the country of Belgium.
Agency product description and boundaries

LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.5

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: LUX-Dx II Plus

3. Open the official notice or contact

Official recall ID
Z-2049-2026
Brand
Boston Scientific
Classification
Class II
Units affected
148 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Saint Paul, MN, United States
Firm address
4100 Hamline Ave N, 55112-5700

Identifying details (verbatim from the notice)

Product codes / lots
LUX-Dx II Plus UDI-DI: 00802526620706 (US), 00802526623707 (EU/AU); SERVER SW LATITUDE DRAGON EU UDI-DI: 00802526613821; SERVER SW LATITUDE DRAGON AU UDI-DI: 00802526613814; Serial Number: 100085 100676 100909 100944 101811 102341 102405 102864 102956 102978 103018 105283 105697 106372 106429 106690 107386 107460 108132 108429 108433 108641 108697 108805 108861 108901 109663 109699 109836 110700 110737 110858 111048 111119 111269 111290 111362 111780 111987 112265 112549 113400 113495 113720 114072 114393 115310 115411 115835 115862 115890 115981 116066 116090 116197 117180 117533 117560 117820 117924 117960 119462 120931 121080 121203 121267 121483 122855 123305 123412 124378 124458 124557 124791 124987 125008 126249 127182 127441 127758 127821 128106 128506 129528 129763 129782 130474 131469 132487 132549 133195 133786 133989 134600 135362 136038 136746 136884 137047 137134 137319 137449 138039 138059 138334 138759 139080 139165 140319 141181 142273 143441 143887 144590 144777 145284 145336 145684 146709 147185 147618 148972 149028 153330 153628 155199 155735 155898 156145 156895 157872 158926 159162 160236 160471 160615 161307 161762 162801 162821 163556 163802 167866 169673 169701 171284 172766 175538
Distribution
Worldwide distribution - US Nationwide and the country of Belgium.
Firm notified the public via
Letter
FDA event ID
98679
Recall initiated
March 30, 2026
FDA classified
May 1, 2026
FDA report date
May 13, 2026