FreshRecalls

FDAClass II · Medical device · Recalled May 13, 2026

Medline Convenience Kits recall — Medline Industries, LP

Hazard: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Class II: may cause temporary or medically reversible health consequences.

Official recall ID
Z-2103-2026
Brand
Medline Industries
Classification
Class II
Units affected
1056 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Description

Medline Convenience Kits: 1) CENTRAL LINE UNIVERSAL INSERTI, Model Number: CVI3780; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43

Affected products

Companies & origin

Recalling firm location
Northfield, IL, United States
Firm address
3 Lakes Dr, 60093-2753

Identifying details (verbatim from the notice)

Product codes / lots
1) CVI3780, UDI-DI: 10193489434866(each), 40193489434867(case), Lot Number: 23KBO637; 2) CVI3780, UDI-DI: 10193489434866(each), 40193489434867(case), Lot Number: 24ABC276; 3) DM1035, UDI-DI: 10193489442755(each), 40193489442756(case), Lot Number: 21BBN221; 4) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 21ABJ003; 5) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 21ABO838; 6) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 21DBD662; 7) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 21IBJ337; 8) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 21JBJ253; 9) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 22CBU342; 10) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 22EBL039; 11) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 22GBF510; 12) DYKMBNDL43, UDI-DI: 10193489478891(each), 40193489478892(case), Lot Number: 22GBN333
Distribution
Worldwide - US Nationwide distribution.
Firm notified the public via
Letter
FDA event ID
98642
Recall initiated
January 7, 2026
FDA classified
May 7, 2026
FDA report date
May 13, 2026

Source: U.S. Food & Drug Administration. This page mirrors the official public-domain record verbatim; the linked government source is authoritative. Source last checked July 24, 2026 at 3:00 AM UTC.

Report a data error on this page