FreshRecalls

FDAClass II · Medical device · Recalled May 13, 2026

Medline Convenience Kits recall — Medline Industries, LP

Hazard: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Class II: may cause temporary or medically reversible health consequences.

Official recall ID
Z-2127-2026
Brand
Medline Industries
Classification
Class II
Units affected
175 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Description

Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501G; 3) ARTHROSCOPY PACK, Model Number: DYNJ32715F; 4) CHS ARTHROSCOPY PACK, Model Number: DYNJ50834C; 5) ARTHROSCOPY TRAY-LF, Model Number: DYNJ51069

Affected products

Companies & origin

Recalling firm location
Northfield, IL, United States
Firm address
3 Lakes Dr, 60093-2753

Identifying details (verbatim from the notice)

Product codes / lots
1) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 21JBJ732; 2) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22ABJ831; 3) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22BBD031; 4) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22FBP788; 5) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22IBE795; 6) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22IBF936; 7) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 23BBJ847; 8) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 23BBL738; 9) DYNJ22501G, UDI-DI: 10889942234110(each), 40889942234111(case), Lot Number: 21ABK153; 10) DYNJ22501G, UDI-DI: 10889942234110(each), 40889942234111(case), Lot Number: 21CBZ902; 11) DYNJ32715F, UDI-DI: 10889942530038(each), 40889942530039(case), Lot Number: 21ADB529; 12) DYNJ50834C, UDI-DI: 10889942568178(each), 40889942568179(case), Lot Number: 21BDA710; 13) DYNJ51069, UDI-DI: 10889942118410(each), 40889942118411(case), Lot Number: 21BDB116
Distribution
Worldwide - US Nationwide distribution.
Firm notified the public via
Letter
FDA event ID
98642
Recall initiated
January 7, 2026
FDA classified
May 7, 2026
FDA report date
May 13, 2026

Source: U.S. Food & Drug Administration. This page mirrors the official public-domain record verbatim; the linked government source is authoritative. Source last checked July 24, 2026 at 3:00 AM UTC.