FDA Class I · Medical device · Recall Z-2160-2026 · Recalled June 3, 2026
RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers recall — Bolton Medical Inc. Compare identifiers Agency action Open official record ↗ Agency report date: June 3, 2026.
Source checks: October 22, 2026 at 6:00 PM UTC (oldest successful check).
Sources, dates and coverage limits We collect recent notices on a six-hour schedule using a rolling 90-day source request. Older retained records may also appear. This is a recall index, not a catalogue of all products or a complete recall history.
FDA: stored recall dates April 29, 2026 to October 14, 2026. USDA FSIS is not aggregated here. Check USDA meat, poultry and processed egg notices . EPA and U.S. Coast Guard are not aggregated. Source delays and missing identifiers can prevent a match.
1. Compare your item This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers Company / brand Bolton Medical Lot / serial / date codes All Lots/(Ref Number/Proximal Diameter/DI)28-N4-32-104-32U/ 32/ (01)00843576150461; 28-N4-32-164-28U/ 32/ (01)00843576150959; 28-N4-32-164-32U/ 32/ (01)00843576150591; 28-N4-32-209-28U/ 32/ (01)00843576151055; 28-N4-32-209-32U/ 32/ (01)00843576150720; 28-N4-32-259-32U/ 32/ (01)00843576150850; 28-N4-34-109-34U/ 34/ (01)00843576150478; 28-N4-34-154-30U/ 34/ (01)00843576150966; 28-N4-34-154-34U/ 34/ (01)00843576150607; 28-N4-34-209-30U/ 34/ (01)00843576151062; 28-N4-34-209-34U/ 34/ (01)00843576150737; 28-N4-34-259-34U/ 34/ (01)00843576150867; 28-N4-36-109-36U/ 36/ (01)00843576150485; 28-N4-36-154-32U/ 36/ (01)00843576150973; 28-N4-36-154-36U/ 36/ (01)00843576150614; 28-N4-36-199-32U/ 36/ (01)00843576151079; 28-N4-36-199-36U/ 36/ (01)00843576150744; 28-N4-36-259-36U/ 36/ (01)00843576150874; 28-N4-38-109-38U/ 38/ (01)00843576150492; 28-N4-38-154-34U/ 38/ (01)00843576150980; 28-N4-38-154-38U/ 38/ (01)00843576150621; 28-N4-38-199-34U/ 38/ (01)00843576151086; 28-N4-38-199-38U/ 38/ (01)00843576150751; 28-N4-38-259-38U/ 38/ (01)00843576150881; 28-N4-40-114-40U/ 40/ (01)00843576150508; 28-N4-40-154-36U/ 40/ (01)00843576150997; 28-N4-40-154-40U/ 40/ (01)00843576150638; 28-N4-40-204-36U/ 40/ (01)00843576151093; 28-N4-40-204-40U/ 40/ (01)00843576150768; 28-N4-40-259-40U/ 40/ (01)00843576150898; 28-N4-42-114-42U/ 42/ (01)00843576150515; 28-N4-42-159-38U/ 42/ (01)00843576151000; 28-N4-42-159-42U/ 42/ (01)00843576150645; 28-N4-42-204-38U/ 42/ (01)00843576151109; 28-N4-42-204-42U/ 42/ (01)00843576150775; 28-N4-42-259-42U/ 42/ (01)00843576150904; 28-N4-44-114-44U/ 44/ (01)00843576150522; 28-N4-44-164-40U/ 44/ (01)00843576151017; 28-N4-44-164-44U/ 44/ (01)00843576150652; 28-N4-44-209-40U/ 44/ (01)00843576151116; 28-N4-44-209-44U/ 44/ (01)00843576150782; 28-N4-44-259-40U/ 44/ (01)00843576151215; 28-N4-44-259-44U/ 44/ (01)00843576150911; 28-N4-46-114-46U/ 46/ (01)00843576150539; 28-N4-46-164-42U/ 46/ (01)00843576151024; 28-N4-46-164-46U/ 46/ (01)00843576150669; 28-N4-46-209-42U/ 46/ (01)00843576151123; 28-N4-46-209-46U/ 46/ (01)00843576150799; 28-N4-46-259-42U/ 46/ (01)00843576151222 Distribution Worldwide - distribution in the US Territory of Puerto Rico and the countries of Argentina, Chile, Canada, Colombia, Mexico, United Kingdom. Agency product description and boundaries RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-34-109-34U 28-N4-34-154-30U 28-N4-34-154-34U 28-N4-34-209-30U 28-N4-34-209-34U 28-N4-34-259-34U 28-N4-36-109-36U 28-N4-36-154-32U 28-N4-36-154-36U 28-N4-36-199-32U 28-N4-36-199-36U 28-N4-36-259-36U 28-N4-38-109-38U 28-N4-38-154-34U 28-N4-38-154-38U 28-N4-38-199-34U 28-N4-38-199-38U 28-N4-38-259-38U 28-N4-40-114-40U 28-N4-40-154-36U 28-N4-40-154-40U 28-N4-40-204-36U 28-N4-40-204-40U 28-N4-40-259-40U 28-N4-42-114-42U 28-N4-42-159-38U 28-N4-42-159-42U 28-N4-42-204-38U 28-N4-42-204-42U 28-N4-42-259-42U 28-N4-44-114-44U 28-N4-44-164-40U 28-N4-44-164-44U 28-N4-44-209-40U 28-N4-44-209-44U 28-N4-44-259-40U 28-N4-44-259-44U 28-N4-46-114-46U 28-N4-46-164-42U 28-N4-46-164-46U 28-N4-46-209-42U 28-N4-46-209-46U 28-N4-46-259-42U
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
2. Agency hazard and consumer action FDA reports: Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.
No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.
Class I: most serious — reasonable probability of serious health consequences or death.
Affected product: RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers recall status
3. Open the official notice or contact View the official FDA enforcement record (openFDA) ↗
Official recall ID Z-2160-2026 Brand All Bolton Medical recalls Classification Class I Recall status Ongoing Initiated Voluntary: Firm initiated Affected products Companies & origin Recalling firm location Sunrise, FL, United States Firm address 799 International Pkwy, 33325-6220 Identifying details (verbatim from the notice) Product codes / lots All Lots/(Ref Number/Proximal Diameter/DI)28-N4-32-104-32U/ 32/ (01)00843576150461; 28-N4-32-164-28U/ 32/ (01)00843576150959; 28-N4-32-164-32U/ 32/ (01)00843576150591; 28-N4-32-209-28U/ 32/ (01)00843576151055; 28-N4-32-209-32U/ 32/ (01)00843576150720; 28-N4-32-259-32U/ 32/ (01)00843576150850; 28-N4-34-109-34U/ 34/ (01)00843576150478; 28-N4-34-154-30U/ 34/ (01)00843576150966; 28-N4-34-154-34U/ 34/ (01)00843576150607; 28-N4-34-209-30U/ 34/ (01)00843576151062; 28-N4-34-209-34U/ 34/ (01)00843576150737; 28-N4-34-259-34U/ 34/ (01)00843576150867; 28-N4-36-109-36U/ 36/ (01)00843576150485; 28-N4-36-154-32U/ 36/ (01)00843576150973; 28-N4-36-154-36U/ 36/ (01)00843576150614; 28-N4-36-199-32U/ 36/ (01)00843576151079; 28-N4-36-199-36U/ 36/ (01)00843576150744; 28-N4-36-259-36U/ 36/ (01)00843576150874; 28-N4-38-109-38U/ 38/ (01)00843576150492; 28-N4-38-154-34U/ 38/ (01)00843576150980; 28-N4-38-154-38U/ 38/ (01)00843576150621; 28-N4-38-199-34U/ 38/ (01)00843576151086; 28-N4-38-199-38U/ 38/ (01)00843576150751; 28-N4-38-259-38U/ 38/ (01)00843576150881; 28-N4-40-114-40U/ 40/ (01)00843576150508; 28-N4-40-154-36U/ 40/ (01)00843576150997; 28-N4-40-154-40U/ 40/ (01)00843576150638; 28-N4-40-204-36U/ 40/ (01)00843576151093; 28-N4-40-204-40U/ 40/ (01)00843576150768; 28-N4-40-259-40U/ 40/ (01)00843576150898; 28-N4-42-114-42U/ 42/ (01)00843576150515; 28-N4-42-159-38U/ 42/ (01)00843576151000; 28-N4-42-159-42U/ 42/ (01)00843576150645; 28-N4-42-204-38U/ 42/ (01)00843576151109; 28-N4-42-204-42U/ 42/ (01)00843576150775; 28-N4-42-259-42U/ 42/ (01)00843576150904; 28-N4-44-114-44U/ 44/ (01)00843576150522; 28-N4-44-164-40U/ 44/ (01)00843576151017; 28-N4-44-164-44U/ 44/ (01)00843576150652; 28-N4-44-209-40U/ 44/ (01)00843576151116; 28-N4-44-209-44U/ 44/ (01)00843576150782; 28-N4-44-259-40U/ 44/ (01)00843576151215; 28-N4-44-259-44U/ 44/ (01)00843576150911; 28-N4-46-114-46U/ 46/ (01)00843576150539; 28-N4-46-164-42U/ 46/ (01)00843576151024; 28-N4-46-164-46U/ 46/ (01)00843576150669; 28-N4-46-209-42U/ 46/ (01)00843576151123; 28-N4-46-209-46U/ 46/ (01)00843576150799; 28-N4-46-259-42U/ 46/ (01)00843576151222 Distribution Worldwide - distribution in the US Territory of Puerto Rico and the countries of Argentina, Chile, Canada, Colombia, Mexico, United Kingdom. FDA event ID 98729 Recall initiated April 22, 2026 FDA classified May 22, 2026 FDA report date June 3, 2026 Other medical device recalls from around the same time Atlan A350XL recall — Draeger, Inc. FDA · June 10, 2026 Atlan A350 recall — Draeger, Inc. FDA · June 10, 2026 Automated Impella Controller (AIC) with the below product descriptions and corresponding P recall — Abiomed, Inc. FDA · June 3, 2026 Medline kits containing Huons Co recall — Medline Industries, LP FDA · June 3, 2026 Medline kits containing Huons Co recall — Medline Industries, LP FDA · June 3, 2026
Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes;
the linked government source is authoritative. Source last checked October 22, 2026 at 6:00 PM UTC. Record retrieved September 1, 2026 at 6:00 PM UTC.