Fresh Recalls

FDAClass II · Medical device · Recall Z-2200-2026 · Recalled May 20, 2026

LVIS Intraluminal Support Device recall — MICROVENTION INC.

Agency report date: May 20, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
LVIS Intraluminal Support Device
Company / brand
MICROVENTION
Lot / serial / date codes
Lots: 0000616462, 0000619851, 0000552987, 0000551725. UDI: (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851,
Distribution
International distribution to the country of China.
Agency product description and boundaries

LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: LVIS Intraluminal Support Device

3. Open the official notice or contact

Official recall ID
Z-2200-2026
Brand
MICROVENTION
Classification
Class II
Units affected
38
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Aliso Viejo, CA, United States
Firm address
35 Enterprise, 92656

Identifying details (verbatim from the notice)

Product codes / lots
Lots: 0000616462, 0000619851, 0000552987, 0000551725. UDI: (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851,
Distribution
International distribution to the country of China.
FDA event ID
98755
Recall initiated
April 6, 2026
FDA classified
May 14, 2026
FDA report date
May 20, 2026
Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked September 7, 2026 at 6:00 AM UTC. Record retrieved August 18, 2026 at 6:00 PM UTC.