FDAClass I · Medical device · Recalled June 10, 2026

Pediatric care bed recall — KAYSERBETTEN GMBH & CO. KG

Hazard: If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).

Class I: most serious — reasonable probability of serious health consequences or death.

Official recall ID
Z-2212-2026
Brand
KAYSERBETTEN & CO. KG
Classification
Class I
Units affected
372 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Description

Pediatric care bed; Product Designation: KayserBett IDA;

Affected products

Companies & origin

Recalling firm location
Bad Fallingbostel, Germany
Firm address
Rieper Str. 12, N/A

Identifying details (verbatim from the notice)

Product codes / lots
Product Designation: KayserBett IDA; UDI-DI: 426038961IDAGU; Serial Number: 386, 387, 388, 389, 390, 391, 392, 635, 636, 637, 638, 639, 641, 642, 643, 644, 645, 646, 647, 1072, 1073, 1074, 1075, 1076, 1077, 1078, 1079, 1080, 1081, 1082, 1083, 1085, 1086, 1087, 1088, 1089, 1090, 1110, 1111, 1446, 1447, 1448, 1449, 1648, 1649, 1950, 1975, 1976, 1982, 1983, 1998, 2001, 2003, 2004, 2337, 2338, 2339, 2340, 2341, 2342, 2343, 2344, 2345, 2396, 2763, 2764, 2765, 2766, 2767, 2770, 2771, 2772, 2773, 2774, 3087, 3088, 3089, 3090, 3091, 3092, 3167, 3299, 3300, 3301, 3302, 3303, 3304, 3305, 3306, 3309, 3310, 3311, 3312, 3313, 3314, 3740, 3741, 3760, 4223, 4232, 4233, 4234, 4235, 4236, 4237, 4239, 4240, 4242, 4243, 4244, 4245, 4246, 4247, 4248, 4249, 4250, 4567, 4568, 4569, 4570, 4571, 4572, 4573, 4574, 4575, 4576, 4577, 4578, 4579, 4580, 4581, 4582, 4583, 4584, 4986, 5238, 5254, 5255, 5256, 5257, 5258, 5259, 5637, 5919, 5920, 5921, 5922, 5923, 5924, 5925, 5926, 5937, 5938, 5939, 5940, 5948, 6332, 6338, 6339, 6340, 6341, 6342, 6343, 6344, 6345, 6346, 6347, 6351, 6355, 6795, 6796, 7176, 7177, 7178, 7179, 7180, 7181, 7182, 7183, 7184, 7185, 7186, 7187, 7188, 7189, 7190, 7191, 7192, 7193, 7194, 7195, 7617, 7618, 7619, 7620, 7621, 7622, 7623, 8242, 8243, 8244, 8245, 8588, 8589, 8590, 8591, 8592, 8593, 8594, 8595, 8596, 8597, 8598, 8599, 8600, 8601, 8602, 8603, 8608, 8609, 8691, 8704, 8912, 8921, 8922, 8923, 8924, 8925, 8926, 8927, 8928, 8929, 8930, 9725, 9726, 9727, 10350, 10351, 10352, 10819, 10820, 10821, 10822, 10823, 10824, 10825, 11186, 11367, 11368, 11369, 11370, 11371, 11372, 11797, 11798, 11799, 11800, 11801, 11802, 11803, 11804, 11805, 11806, 11807, 11808, 12247, 12248, 12249, 12250, 12251, 12252, 12253, 12254, 12255, 12256, 12257, 12397, 12398, 12399, 12782, 12783, 12963, 12964, 12965, 12966, 12967, 12968, 12969, 12970, 12971, 12972, 13486, 13487, 13488, 13489, 13490, 13491, 13492, 13493, 13494, 13495, 13496, 13497, 13912, 13913, 13914, 13915, 13916, 13917, 13918, 14942, 14944, 14952, 14953, 15912, 15913, 15914, 16403, 16404, 16405, 16406, 16407, 16408, 16409, 16410, 16411, 16984, 16985, 16986, 16987, 16988, 16989, 16990, 16991, 16992, 16993, 16994, 17577, 17578, 17579, 17580, 17581, 17582, 17583, 17584, 17585, 17586, 17587, 18494, 19112, 19113, 19114, 20151, 20152, 20153, 20154, 20155, 20156, 20157, 20158, 20963, 20964, 20965, 20966, 20967, 20968, 20969, 20970, 20972, 21192, 21193, 21194;
Distribution
Distribution US Nationwide and Canada.
Firm notified the public via
Letter
FDA event ID
98820
Recall initiated
May 4, 2026
FDA classified
June 4, 2026
FDA report date
June 10, 2026