Fresh Recalls

FDAClass II · Medical device · Recall Z-2355-2026 · Recalled June 17, 2026

NeoSpan¿ SuperElastic Compression Staple w/Instruments Catalog Number and Product Name T50 recall — In2bones USA, LLC

Agency report date: June 17, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
NeoSpan¿ SuperElastic Compression Staple w/Instruments Catalog Number and Product Name T50
Company / brand
In2bones USA
Lot / serial / date codes
Catalog No. Lot Code T50 SN010 1962242 1964665 2054541 2039361 T50 SN012 1963511 T50 SN020 1971252 1990642 2027441 T50 SN110 1971241 T50 SN115 19647210 T50 SN118 1990631
Distribution
U.S.: AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, LA, MD, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, and WI O.U.S.: Belgium, Canada, Colombia, France, Ireland, Netherlands, Spain, United Kingdom,
Agency product description and boundaries

NeoSpan¿ SuperElastic Compression Staple w/Instruments Catalog Number and Product Name T50 SN010 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 10x10x10 UDI-DI Code: 00817906020083 T50 SN012 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 12x12x12 UDI-DI Code: 00817906020090 T50 SN020 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 20x20x20 UDI-DI Code: 00817906020151 T50 SN110 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 10x15x13 UDI-DI Code: 00817906020113 T50 SN115 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 15Wx15x15 UDI-DI Code: 00817906020137 T50 SN118 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 18x14x14 UDI-DI Code: 00810021861364 Single/multiple component metallic bone fixation appliances and accessories The In2Bones USA NeoSpan" Compression Staple Implant w/instruments is indicated for hand and foot bone fragments, osteotomy fixation and joint arthrodesis.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Due to the Magnetic Resonance Conditional Symbol missing on the carton, inner tray, and patient label.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: NeoSpan¿ SuperElastic Compression Staple w/Instruments Catalog Number and Product Name T50 recall status

3. Open the official notice or contact

Official recall ID
Z-2355-2026
Brand
All In2bones USA recalls
Classification
Class II
Units affected
306 kits
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Cordova, TN, United States
Firm address
7750a Trinity Rd, N/A, 38018-2722

Identifying details (verbatim from the notice)

Product codes / lots
Catalog No. Lot Code T50 SN010 1962242 1964665 2054541 2039361 T50 SN012 1963511 T50 SN020 1971252 1990642 2027441 T50 SN110 1971241 T50 SN115 19647210 T50 SN118 1990631
Distribution
U.S.: AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, LA, MD, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, and WI O.U.S.: Belgium, Canada, Colombia, France, Ireland, Netherlands, Spain, United Kingdom,
Firm notified the public via
N/A
FDA event ID
98978
Recall initiated
May 11, 2026
FDA classified
June 8, 2026
FDA report date
June 17, 2026

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 6:00 PM UTC. Record retrieved September 15, 2026 at 6:00 PM UTC.

Report a data error on this page