FDAClass I · Medical device · Recalled June 24, 2026
Description/REF recall — ARROW INTERNATIONAL, LLC
Hazard: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Class I: most serious — reasonable probability of serious health consequences or death.
View the official FDA enforcement record (openFDA) ↗
- Official recall ID
- Z-2388-2026
- Brand
- ARROW INTERNATIONAL
- Classification
- Class I
- Units affected
- 62,719
- Recall status
- Ongoing
- Initiated
- Voluntary: Firm initiated
Description
Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 CM/ASK-15123-UOC1, AHDC KIT: 3-L 12 FR X 20 CM/ASK-15123-UPM, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-P1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XP1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPCN1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-P1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPCN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPN1A, AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XPCN1A, HEMODIALYSIS KIT 2L: 12 FR X 20CM AGB/AK-25122-CDC, HEMODIALYSIS KIT: 2-LUMEN 14 FR X 20 CM/AK-25142-F, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM1, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM2, AHDC KIT: 2L 14 FR X 15 CM/ASK-22142-UOC1, AHDC AGB KIT: 2-L 14 FR X 15 CM/ASK-22142-UPM, HEMODIALYSIS KIT: 2L 14 FR X 20 CM/ASK-25142-UNC1, AHDC KIT: 2L 14 FR X 20 CM/ASK-25142-UOC1, AHDC AGB KIT: 2L 14 FR X 20 CM/ASK-25142-UPM, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/ASK-26142-UNC1, HEMODIALYSIS KIT: 2LUMEN 12 FR X 16 CM/AU-22122-F, HEMODIALYSIS KIT: 2L 12 FR X 16 CM AGB/CDC-22122-1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A-U, HEMODIALYSIS KIT 2L: 14 FR X 15 CM AGB/CDC-22142-1A, HEMODIALYSIS KIT: 2L 14 FR X 15 CM AGB/CDC-22142-XC1A, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A-C, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XCN1A-U, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XN1A-U, HEMODIALYSIS KIT: 2L 12 FR X 13 CM AGB/CDC-23122-XU1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-X1A, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XN1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-XU1A, HEMODIALYSIS KIT 2L: 14 FR X 20 CM AGB/CDC-25142-1A, HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB/CDC-25142-XC1A, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A-C, HEMODIALYSIS KIT: 2L 12 FR X 25 CM AGB/CDC-26122-1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XCN1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XN1A, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/CDC-26142-1A, HEMODIALYSIS KIT: 2L 14 FR X 25 CM AGB/CDC-26142-X1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XCN1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XN1A, HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-25122-F, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 16 CM/CDC-22123-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 20 CM/CDC-25123-XCN1A, LBCVC AGB KIT:3-L 12 FR X 20 CM/CDC-25123-XN1A
Affected products
Companies & origin
- Recalling firm location
- Morrisville, NC, United States
- Firm address
- 3015 Carrington Mill Blvd, 27560-5437
Identifying details (verbatim from the notice)
- Product codes / lots
- REF(Material)/UDI-DI/Lot(Batch): AK-15122-F/10801902197136, 10801902193275, 10801902168525/33F24E0348, 33F24J0280, 33F24K0119, 33F24L0601, 33F25A0477, 33F25C0908, 33F24A0850; ASK-12123-UOC1/10801902193275/33F25B0430, 33F25D0119; ASK-12123-UPM/10801902193275, 10801902222463/33F25E0535, 33F25G0188, 33F25G0189, 33F26C0333, 33F25D0262; ASK-15123-UOC1/10801902193275/33F25C0378, 33F25D0121; ASK-15123-UPM/10801902193275, 10801902159950/33F25E0718, 33F25H0447, 33F26A1389, 33F24E0649; CDC-12123-P1A/40801902111768, 10801902193275/33F24D0632, 33F24F0758, 33F24L0465, 33F25A0281, 33F25E0794; CDC-12123-XP1A/10801902193275, 30801902111785, 10801902213416/33F25J0723, 33F26A1382, 33F26B0433, 36F25A0007, 36F25B0009, 36F25E0018, 33F24D0299; CDC-12123-XPCN1A/10801902213416, 10801902193275/33F24D0950, 33F24K0417, 33F24L0466, 33F25A0216, 33F25B0667, 33F25D0436, 33F25D0666, 33F25E0693, 33F25G0239, 33F25J0247, 33F26A1556, 33F26B0801, 33F26C0767, 33F25B0668; CDC-12123-XPN1A/10801902193275, 10801902213409/33F25D0554, 33F25G0240, 33F25J0312, 33F26B0214,33F26C0199, 33F25A0181; CDC-15123-P1A/10801902193275/33F25D0437, 33F25H0547,33F25B0613; CDC-15123-XPCN1A/10801902213430, 10801902193275/33F24F0107, 33F24F0998, 33F24K0403, 33F24L0834, 33F25A0283, 33F25B0527, 33F25E0650, 33F25E0745, 33F25F0085, 33F25H0596, 33F25J0314, 33F26A1464, 33F26B0883, 33F25B0669; CDC-15123-XPN1A/0801902193275, 10801902213423, 10801902116793/33F25D0439, 33F25G0238, 33F25H0597, 33F25J0315, 33F26A1468, 33F26B0826, 33F24C0461; CDC-16123-XPCN1A/10801902193275/33F25B0671, 33F25D0440, 33F25E0651, 33F25G0236, 33F25H0529, 33F25J0316, 33F25G0237, AK-25122-CDC/10801902193275/33F24G0370, 33F24K0135, 33F25C0120, 33F25C0969, 33F25E0942, 33F24G0371; AK-25142-F/10801902193275, 70801902123205, 10801902167795/33F25C0863, 33F25E0087, 33F25H0038, 33F25J0846, 33F26B0531, 33F24D0089; ASK-22122-MM1/10801902193275/33F25E0546; ASK-22122-MM2/10801902193275, 10801902221923, 10801902144956/33F25G0172, 33F25L0031, 33F25M0310, 33F26A1652, 33F26B0151, 33F26B0450, 33F24F0854; ASK-22142-UOC1/10801902193275/33F25C0545, 33F25D0122; ASK-22142-UPM/10801902193275, 10801902222487, 10801902202281/33F25D0594, 33F25F0116, 33F25G0187, 33F26A1390, 33F26B0205, 33F24D0396; ASK-25142-UNC1/10801902193275/33F25C0655, 33F24K0561; ASK-25142-UOC1/10801902193275/33F25B0362, 33F25D0123; ASK-25142-UPM/10801902193275/33F25E0595, 33F25H0446, 33F25J0381, 33F26A1446, 33F25J0092; ASK-26142-UNC1/10801902193275/33F25B0427, 33F25K0058; AU-22122-F/10801902193275/33F24K0269, 33F25A0584, 33F25C0862, 33F25E0245, 33F25J0567; CDC-22122-1A/20801902123484, 10801902193275/, CDC-22122-XCN1A-U/33F24F0996, 33F24L0374, 33F25C0461; CDC-22122-XN1A/10801902193275/33F25C0239, 33F25E0875, 33F25J0419; CDC-22122-XN1A-U/10801902193275, 10801902213621/33F25A0309, 33F25E0876, 33F26B0217; CDC-22142-1A/50801902123515, 10801902193275/33F24D0928, 33F24F0103, 33F24F0896, 33F24L0111, 33F24L0955, 33F25B0500, 33F25D0447, 33F25E0654, 33F25G0278, 33F25H0577, 33F25J0319; CDC-22142-XC1A/10801902193275/33F25A0284, 33F25J0180; CDC-22142-XCN1A/10801902193275, 10801902213539/33F25A0285, 33F25B0529, 33F25D0449, 33F25E0655, 33F25E0818, 33F25F0214, 33F25G0233, 33F26A1477, 33F26B0898, 33R25J0455; CDC-22142-XCN1A-C/10801902193275/33F24L0473, 33F25B0615, 33F25F0213; CDC-22142-XN1A/10801902193275, 10801902213522/33F25C0240, 33F25E0578, 33F25F0215, 33F25H0528, 33F25J0321, 33F26A1383, 33F26B0215, 33F26C0016; CDC-22142-XN1A-C/10801902193275/33F25B0485, 33F25E0879; CDC-23122-XCN1A-U/10801902193275/33F24L0279, 33F25D0492; CDC-23122-XN1A-U/10801902193275/33F25E0579; CDC-23122-XU1A/10801902193275/33F25D0493, 33F25H0432, 33F25J0248; CDC-25122-1A/10801902193275/33F25B0619, 33F25E0881; CDC-25122-X1A/10801902193275/33F25E0882, 33F25J0422; CDC-25122-XCN1A/10801902213492, 10801902193275/33F24F0929, 33F25C0242, 33F25D0498, 33F25E0883, 33F25G0234, 33F26B0439, 33F26C0214; CDC-25122-XCN1A-U/10801902193275/33F24L1041, 33F25B0620, 33F25F0121, 33F25G0226; CDC-25122-XN1A/10801902193275, 10801902213485/33F25C0243, 33F25E0884, 33F25K0183, 33R25E0884; CDC-25122-XU1A/10801902193275/33F25B0621, 33F25E0886, 33F25H0568; CDC-25142-1A/50801902123584, 10801902193275/33F24F0837, 33F24K0457, 33F24L0113, 33F24L1031, 33F25B0622, 33F25D0503, 33F25E0889, 33F25F0106, 33F25J0127; CDC-25142-XC1A/10801902193275/33F25B0467; CDC-25142-XCN1A/10801902193275, 10801902213553/33F25A0312, 33F25E0657, 33F25E0821, 33F25G0231, 33F25H0527, 33F25J0428, 33F26B0452, 33F26C0133; CDC-25142-XCN1A-C/10801902193275/33F24L0377, 33F25D0238; CDC-25142-XN1A/10801902193275, 10801902213546/33F25B0624, 33F25D0504, 33F25E0823, 33F25G0232, 33F26B0437; CDC-25142-XN1A-C/10801902193275/33F24L0485, 33F25B0676; CDC-26122-1A/60801902123611, 10801902193275/33F24F0084, 33F24K0475, 33F25G0389; CDC-26122-XCN1A/10801902213515, 10801902193275/33F24F1028, 33F24L0117, 33F25D0308, 33F25D0557, 33F25F0219; CDC-26122-XN1A/10801902193275/33F25E0451; CDC-26142-1A/10801902193275/33F24L0488, 33F24L1025, 33F25C0368, 33F25E0891, 33F25H0406, 33F25J0116; CDC-26142-X1A/10801902193275/33F25D0558, 33F25E0892; CDC-26142-XCN1A/10801902193275, 10801902213577/33F24L0941, 33F25B0530, 33F25D0507, 33F25E0699, 33F25H0525, 33F26B0396, 33F26B0895, 33F26C0780; CDC-26142-XN1A/10801902193275, 10801902213560/33F24L0927, 33F25E0580, 33F25H0526, 33F26B0393; AK-25122-F/ CDC-22122-XCN1A/10801902193275/33F24G0371, 33F24L0742, 33F25C0822, 33F25E0084, 33F25J0483, 33F25L0400; CDC-22123-XCN1A/10801902213591, 10801902193275/33F24D0842,33F24F1018, 33F24L0110, 33F25C0205, 33F25D0613, 33F25E0878, 33F26B0455; CDC-25123-XCN1A/10801902213614, 10801902193275/33F24D0818, 33F24L0483, 33F25C0244, 33F25D0502, 33F25E0888, 33F25G0228, 33F25J0425, 33F26B0395, 33F26B0894; CDC-25123-XN1A/10801902193275/33F25C0332, 33F25G0229
- Distribution
- US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
- Firm notified the public via
- Letter
- FDA event ID
- 98793
- Recall initiated
- April 21, 2026
- FDA classified
- June 12, 2026
- FDA report date
- June 24, 2026
Source: U.S. Food & Drug Administration. This page mirrors the official public-domain record verbatim;