FreshRecalls

FDAClass II · Medical device · Recalled June 24, 2026

Medline convenience kits recall — Medline Industries, LP

Hazard: Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Class II: may cause temporary or medically reversible health consequences.

Official recall ID
Z-2447-2026
Brand
Medline Industries
Classification
Class II
Units affected
49,654 kits total
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Description

Medline convenience kits: HYSTERECTOMY PACK-LF DYNJ20485L

Affected products

Companies & origin

Recalling firm location
Northfield, IL, United States
Firm address
3 Lakes Dr, 60093-2753

Identifying details (verbatim from the notice)

Product codes / lots
UDI-DI 10195327024369 Lot 25LMJ031
Distribution
US Nationwide distribution. OUS distribution pending.
Firm notified the public via
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
FDA event ID
98951
Recall initiated
April 27, 2026
FDA classified
June 12, 2026
FDA report date
June 24, 2026

Source: U.S. Food & Drug Administration. This page mirrors the official public-domain record verbatim; the linked government source is authoritative. Source last checked July 24, 2026 at 3:00 AM UTC.

Report a data error on this page