FreshRecalls

⚠ Data freshness notice: the most recent refresh of NHTSA was incomplete, so a very recent recall may be missing. Everything shown is accurate as of the last checked time on this page — for anything time-critical, confirm with the official sources linked below.

FDAClass II · Medical device · Recalled June 24, 2026

MEDLINE Medical Procedure Kits labeled as recall — Medline Industries, LP

Hazard: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Class II: may cause temporary or medically reversible health consequences.

Official recall ID
Z-2485-2026
Brand
Medline Industries
Classification
Class II
Units affected
72 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Description

MEDLINE Medical Procedure Kits labeled as: LAPAROSCOPY COMBINED - NO NDL, Medline Kit Number/SKU DYNJ37378A

Affected products

Companies & origin

Recalling firm location
Northfield, IL, United States
Firm address
3 Lakes Dr, 60093-2753

Identifying details (verbatim from the notice)

Product codes / lots
UDI/DI 10889942113583 (each), 40889942113584 (case); Lot Numbers: DYNJ37378A, DYNJ37378A, DYNJ37378A.
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Firm notified the public via
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
FDA event ID
98994
Recall initiated
May 4, 2026
FDA classified
June 17, 2026
FDA report date
June 24, 2026

Source: U.S. Food & Drug Administration. This page mirrors the official public-domain record verbatim; the linked government source is authoritative. Source last checked July 23, 2026 at 11:04 PM UTC.

Report a data error on this page