Fresh Recalls

FDAClass II · Medical device · Recall Z-2594-2026 · Recalled July 1, 2026

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes recall — Medline Industries, LP

Agency report date: July 1, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes
Company / brand
Medline Industries
Lot / serial / date codes
DYNJ41089 UDI-DI 10884389891360 Lot 21BBB592 DYNJ62827A UDI-DI 10193489306606 Lot 24BBN289 DYNJ47899A UDI-DI 10889942574681 Lots 21GBI377 21IBL299 21LBE099 21LBU529 22JBV803 DYNJ44123B UDI-DI 10195327381653 Lots 23FBR833 23GBN041 23HBW014 23JBO264 23JBW218 23KBI873 DYNJ68037A UDI-DI 10195327003630 Lots 21JBQ304 21LBQ205 22ABN185 22DBT349 22EBL611 22HBH550 22IBN880 22JBH689 22KBS987 22OBA368 23ABD642 23BBH830 23DBL865 23EBT367 23FBM946 23GBT261 DYNJ0101291F UDI-DI 10195327286644 Lots 22KDB751 23EDA119 23EDB388 23FDA396 23GDA313 23HDA759 23LDB909 24ADB667 24GDA294 24HDB515
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
Agency product description and boundaries

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. CENTRAL LINE TRAY DYNJ41089 FISTULAGRAM PACK DYNJ62827A FISTULOGRAM PACK DYNJ47899A INVASIVE LINE INSERTION DYNJ44123B NON VASCULAR PACK DYNJ68037A VASCULAR ACCESS PACK-LF DYNJ0101291F

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes

3. Open the official notice or contact

Official recall ID
Z-2594-2026
Brand
Medline Industries
Classification
Class II
Units affected
113, 843 kits
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Northfield, IL, United States
Firm address
3 Lakes Dr, N/A, 60093-2753

Identifying details (verbatim from the notice)

Product codes / lots
DYNJ41089 UDI-DI 10884389891360 Lot 21BBB592 DYNJ62827A UDI-DI 10193489306606 Lot 24BBN289 DYNJ47899A UDI-DI 10889942574681 Lots 21GBI377 21IBL299 21LBE099 21LBU529 22JBV803 DYNJ44123B UDI-DI 10195327381653 Lots 23FBR833 23GBN041 23HBW014 23JBO264 23JBW218 23KBI873 DYNJ68037A UDI-DI 10195327003630 Lots 21JBQ304 21LBQ205 22ABN185 22DBT349 22EBL611 22HBH550 22IBN880 22JBH689 22KBS987 22OBA368 23ABD642 23BBH830 23DBL865 23EBT367 23FBM946 23GBT261 DYNJ0101291F UDI-DI 10195327286644 Lots 22KDB751 23EDA119 23EDB388 23FDA396 23GDA313 23HDA759 23LDB909 24ADB667 24GDA294 24HDB515
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
Firm notified the public via
Letter
FDA event ID
99062
Recall initiated
February 27, 2026
FDA classified
June 25, 2026
FDA report date
July 1, 2026
Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked September 7, 2026 at 6:00 AM UTC. Record retrieved September 7, 2026 at 6:00 AM UTC.

Report a data error on this page