Fresh Recalls

FDAClass I · Medical device · Recall Z-2605-2026 · Recalled July 8, 2026

Omnipod (Eros) Pod recall — Insulet Corporation

Agency report date: July 8, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Omnipod (Eros) Pod
Company / brand
Insulet
Lot / serial / date codes
1. Model/REF Number: POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72514. 2. Model/REF Number: ZXR425 (10-Pack), ZXR420 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: L72516, L72521. 3. Model/REF Number: ZXP425 (10 Pack), ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72515, L72509, L72513, L72511, L72510, L71480, L72514.
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.
Agency product description and boundaries

Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class I: most serious — reasonable probability of serious health consequences or death.

Affected product: Omnipod (Eros) Pod recall status

3. Open the official notice or contact

Official recall ID
Z-2605-2026
Brand
All Insulet recalls
Classification
Class I
Units affected
289,046 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Acton, MA, United States
Firm address
50 & 100 Nagog Park, N/A, 01720-3440

Identifying details (verbatim from the notice)

Product codes / lots
1. Model/REF Number: POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72514. 2. Model/REF Number: ZXR425 (10-Pack), ZXR420 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: L72516, L72521. 3. Model/REF Number: ZXP425 (10 Pack), ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72515, L72509, L72513, L72511, L72510, L71480, L72514.
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.
Firm notified the public via
Press Release
FDA event ID
99034
Recall initiated
May 20, 2026
FDA classified
June 30, 2026
FDA report date
July 8, 2026

More Insulet recalls

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 6:00 PM UTC. Record retrieved October 6, 2026 at 6:00 PM UTC.

Report a data error on this page