Fresh Recalls

FDAClass II · Medical device · Recall Z-2607-2026 · Recalled July 8, 2026

Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G recall — Medline Industries, LP

Agency report date: July 8, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G
Company / brand
Medline Industries
Lot / serial / date codes
DYNJ901328G UDI-DI 10198459601538 Lots 26ABD786 26CBD768 26DBJ434 DYNJ903465K UDI-DI 10198459292743 Lots 25GDB257 26ADA464 26BDA587 26DDA021 DYNJ909090G UDI-DI 10198459406027 Lots 25LBH690 25LBP697 DYNJ906102M UDI-DI 10198459406409 Lots 25JBE315 25KBI700 25KBM827 25LBB028 26ABH106 26CBQ206 DYNJ902664M UDI-DI 10198459349126 Lots 25ILA177 25KLA500 DYNJ902664N UDI-DI 10198459675287 Lot 26BLA783
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Agency product description and boundaries

Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G HEART A DYNJ903465K OPEN HEART DYNJ909090G SILVER CROSS OPEN HEART DYNJ906102M VPH OPEN HEART DYNJ902664M DYNJ902664N

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G

3. Open the official notice or contact

Official recall ID
Z-2607-2026
Brand
Medline Industries
Classification
Class II
Units affected
9,272 kits
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Northfield, IL, United States
Firm address
3 Lakes Dr, N/A, 60093-2753

Identifying details (verbatim from the notice)

Product codes / lots
DYNJ901328G UDI-DI 10198459601538 Lots 26ABD786 26CBD768 26DBJ434 DYNJ903465K UDI-DI 10198459292743 Lots 25GDB257 26ADA464 26BDA587 26DDA021 DYNJ909090G UDI-DI 10198459406027 Lots 25LBH690 25LBP697 DYNJ906102M UDI-DI 10198459406409 Lots 25JBE315 25KBI700 25KBM827 25LBB028 26ABH106 26CBQ206 DYNJ902664M UDI-DI 10198459349126 Lots 25ILA177 25KLA500 DYNJ902664N UDI-DI 10198459675287 Lot 26BLA783
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Firm notified the public via
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
FDA event ID
99209
Recall initiated
May 28, 2026
FDA classified
June 26, 2026
FDA report date
July 8, 2026
Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked September 7, 2026 at 6:00 AM UTC. Record retrieved September 7, 2026 at 6:00 AM UTC.

Report a data error on this page