FDAClass II · Medical device · Recalled July 8, 2026
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer recall — Accuray Incorporated
Hazard: In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
Class II: may cause temporary or medically reversible health consequences.
View the official FDA enforcement record (openFDA) ↗
Description
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Affected products
Companies & origin
Identifying details (verbatim from the notice)
Source: U.S. Food & Drug Administration. This page mirrors the official public-domain record verbatim; the linked government source is authoritative. Source last checked July 23, 2026 at 11:04 PM UTC.