⚠ Data freshness notice: the most recent refresh of NHTSA was incomplete, so a
very recent recall may be missing. Everything shown is accurate as of the last checked time on this page —
for anything time-critical, confirm with the official sources linked below.
FDAClass II · Medical device · Recalled July 8, 2026
Covidien Shiley Disposable Decannulation Plug REF DDCP UDI-DI code recall — Covidien LLC
Hazard: Due to product labeling incorrectly including the symbol indicating non-DEHP (Bis(2-ethylhexyl) phthalate as the product includes 30 to 35% DEHP.
Class II: may cause temporary or medically reversible health consequences.
View the official FDA enforcement record (openFDA) ↗
- Official recall ID
- Z-2653-2026
- Brand
- Covidien
- Classification
- Class II
- Units affected
- 293,659 devices
- Recall status
- Ongoing
- Initiated
- Voluntary: Firm initiated
Description
Covidien Shiley Disposable Decannulation Plug REF DDCP UDI-DI code: 10884522004060 Product Codes that include the Covidien Shiley Disposable Decannulation Plug are: Product Number / UDI-DI code 4DCFN 10884522006576 4DCFS 10884522006583 4DFEN 10884522006606 6DCFN 10884522006613 6DCFS 10884522006620 6DFEN 10884522006644 8DCFN 10884522006651 8DCFS 40884522006690 8DFEN 10884522006675 10DCFN 20884522006535 10DCFS 10884522006545 10DFEN 10884522006569 The Shiley Disposable Decannulation Plug (DDCP) is a universal size and fits any size DCFS, DCFN, DFEN tracheostomy tube. The red DDCP snaplock connector is used to occlude the proximal end of the outer cannula on the Shiley Tracheostomy Tubes. Refer to Table 1 for model types. The DDCP blocks airflow through the tube and directs breathing through the mouth and nose. The Shiley Tracheostomy Tube and the accessory products are intended for use in providing tracheal access for airway management.
Affected products
Companies & origin
- Recalling firm location
- Boulder, CO, United States
- Firm address
- 6135 Gunbarrel Ave, N/A, 80301-3214
Identifying details (verbatim from the notice)
- Product codes / lots
- Product Number: DDCP UDI-DI code: 10884522004060 The below product numbers and UDI-DI codes that DDCP are: 4DCFN / 10884522006576 4DCFS / 10884522006583 4DFEN / 10884522006606 6DCFN / 10884522006613 6DCFS / 10884522006620 6DFEN / 10884522006644 8DCFN / 10884522006651 8DCFS / 40884522006690 8DFEN / 10884522006675 10DCFN / 20884522006535 10DCFS / 10884522006545 10DFEN / 10884522006569
- Distribution
- U.S. Nationwide distribution including in the states and territories of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Virgin Islands. The countries of Albania, Andorra, Argentina, Australia, Austria, Belgium, Bhutan, Canada, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Finland, France, Georgia, Germany, Greece, Guadeloupe, Guyana, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Latvia, Lithuania, Luxembourg, Martinique, Mexico, Netherlands, New Caledonia, New Zealand, North Macedonia, Norway, Oman, Paraguay, Peru, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Togo, Trinidad And Tobago, Turkey, Uganda, United Arab Emirates, Uruguay.
- Firm notified the public via
- Letter
- FDA event ID
- 98819
- Recall initiated
- April 23, 2026
- FDA classified
- July 2, 2026
- FDA report date
- July 8, 2026
Source: U.S. Food & Drug Administration. This page mirrors the official public-domain record verbatim;
the linked government source is authoritative. Source last checked July 23, 2026 at 11:04 PM UTC.
Report a data error on this page