FDAClass II · Medical device · Recalled July 15, 2026
Brand Name recall — Elekta, Inc.
Hazard: Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.
Class II: may cause temporary or medically reversible health consequences.
View the official FDA enforcement record (openFDA) ↗
- Official recall ID
- Z-2711-2026
- Brand
- Elekta
- Classification
- Class II
- Units affected
- 586
- Recall status
- Ongoing
- Initiated
- Voluntary: Firm initiated
Description
Brand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: The intended use of the Advanced Gynecological Applicator is intracavitary brachytherapy for cancer treatment of the cervix and endometrium. Optional needles can be placed for interstitial brachytherapy.
Affected products
Companies & origin
- Recalling firm location
- Atlanta, GA, United States
- Firm address
- 400 Perimeter Center Ter NE Ste 50, N/A, 30346-1227
Identifying details (verbatim from the notice)
- Product codes / lots
- Lot Code: 08717213055001, 08717213055018, 08717213055025, 08717213055049, 08717213055056, 08717213055063, 08717213055087, 08717213055094, 08717213055100, 08717213055117, 08717213055131
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TX, UT, VA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran (Islam Republic of), Iraq, Ireland, Italy, Japan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zambia.
- Firm notified the public via
- Letter
- FDA event ID
- 98704
- Recall initiated
- April 3, 2026
- FDA classified
- July 9, 2026
- FDA report date
- July 15, 2026
Source: U.S. Food & Drug Administration. This page mirrors the official public-domain record verbatim;
the linked government source is authoritative. Source last checked July 23, 2026 at 10:00 PM UTC.