Fresh Recalls

FDAClass II · Medical device · Recall Z-2771-2026 · Recalled July 29, 2026

Boston Scientific TheraSphere 360 Y-90 Management Platform recall — Boston Scientific Corporation

Agency report date: July 29, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Boston Scientific TheraSphere 360 Y-90 Management Platform
Company / brand
Boston Scientific
Lot / serial / date codes
UDI/DI 00191506068264, All released software versions (V1.1.2, V1.2.1, V1.2.2. V2.0.1, V2.0.2)
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.
Agency product description and boundaries

Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Boston Scientific TheraSphere 360 Y-90 Management Platform

3. Open the official notice or contact

Official recall ID
Z-2771-2026
Brand
Boston Scientific
Classification
Class II
Units affected
68 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Maple Grove, MN, United States
Firm address
1 Scimed Pl, N/A, 55311-1565

Identifying details (verbatim from the notice)

Product codes / lots
UDI/DI 00191506068264, All released software versions (V1.1.2, V1.2.1, V1.2.2. V2.0.1, V2.0.2)
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.
Firm notified the public via
Telephone
FDA event ID
99236
Recall initiated
June 19, 2026
FDA classified
July 22, 2026
FDA report date
July 29, 2026
Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked September 7, 2026 at 6:00 AM UTC. Record retrieved September 7, 2026 at 6:00 AM UTC.

Report a data error on this page