Fresh Recalls

FDAClass II · Medical device · Recall Z-2821-2026 · Recalled August 5, 2026

GEM Premier 5000 with IQM2 Model/Catalog Number recall — Instrumentation Laboratory

Agency report date: August 5, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
GEM Premier 5000 with IQM2 Model/Catalog Number
Company / brand
Instrumentation Laboratory
Lot / serial / date codes
GEM Premier 5000 Part number: 00024019255, UDI: 08426950870227 GEM Premier 7000 Part number: 00000015279, UDI: 0195226000924 Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
Distribution
US Nationwide distribution.
Agency product description and boundaries

GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: GEM Premier 5000 with IQM2 Model/Catalog Number

3. Open the official notice or contact

Official recall ID
Z-2821-2026
Brand
Instrumentation Laboratory
Classification
Class II
Units affected
122
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Bedford, MA, United States
Firm address
180 Hartwell Rd, N/A, 01730-2443

Identifying details (verbatim from the notice)

Product codes / lots
GEM Premier 5000 Part number: 00024019255, UDI: 08426950870227 GEM Premier 7000 Part number: 00000015279, UDI: 0195226000924 Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
Distribution
US Nationwide distribution.
Firm notified the public via
Letter
FDA event ID
99115
Recall initiated
May 8, 2026
FDA classified
July 29, 2026
FDA report date
August 5, 2026
Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked September 7, 2026 at 6:00 AM UTC. Record retrieved September 7, 2026 at 6:00 AM UTC.

Report a data error on this page