Fresh Recalls

FDAClass II · Medical device · Recall Z-2861-2026 · Recalled August 12, 2026

Brand Name recall — Nuclein LLC

Agency report date: August 12, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Brand Name
Company / brand
Nuclein
Lot / serial / date codes
Lot Code: SN: C01250010; UDI: (01)00850063063005(11)250218(21)C01250010; SW: 6.0.2 SN: C02250006; UDI:(01)00850063063005(11)250225(21)C02250006; SW: 6.0.2
Distribution
US Nationwide distribution in the state of Texas.
Agency product description and boundaries

Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH¿ Rapid PCR Instrument (DASH Instrument) is an automated, fluorescence-based, reverse-transcription polymerase chain reaction (RT-PCR) instrument that extracts specimens and performs nucleic acid amplified tests (NAAT) in DASH test cartridges for the qualitative detection of infectious diseases. The DASH Instrument is intended for use in conjunction with DASH Rapid PCR System tests to aid in the diagnosis of infectious diseases. Component: No.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Due to a software issue which may result in false co-infection (e.g., Positive for Flu A, when SARS-CoV-2 is present).

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Brand Name

3. Open the official notice or contact

Official recall ID
Z-2861-2026
Brand
Nuclein
Classification
Class II
Units affected
2
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Austin, TX, United States
Firm address
8305 Cross Park Dr, 78754-5154

Identifying details (verbatim from the notice)

Product codes / lots
Lot Code: SN: C01250010; UDI: (01)00850063063005(11)250218(21)C01250010; SW: 6.0.2 SN: C02250006; UDI:(01)00850063063005(11)250225(21)C02250006; SW: 6.0.2
Distribution
US Nationwide distribution in the state of Texas.
Firm notified the public via
Letter
FDA event ID
99360
Recall initiated
June 25, 2025
FDA classified
August 3, 2026
FDA report date
August 12, 2026
Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked September 7, 2026 at 6:00 AM UTC. Record retrieved September 7, 2026 at 6:00 AM UTC.

Report a data error on this page