Fresh Recalls

FDAClass II · Medical device · Recall Z-2958-2026 · Recalled August 26, 2026

Arrow-LOK Implant recall — Arrowhead DE LLC

Agency report date: August 26, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Arrow-LOK Implant
Company / brand
Arrowhead DE
Lot / serial / date codes
REF/UDI-DI/Lot: 21-253519/00855967006035/24662119, 242162; 21-253522/00855967006042/24662122, 242163, 220702; 21-253525/00855967006059/24662125, 220703, 242164; 21-253528/00855967006066/24662128, 242165, 230704; 21-253530/00855967006073/242166, 230705, 220705; 21-253533/00855967006080/24662133, 242167, 230706; 21-253535/00855967006097/24662135, 242168, 230707; 21-253540/00855967006110/242169; 21-253545/00855967006127/242170; 21-253550/00855967006134/24662150; 21-353513/00855967006004/24662113, 242160; 21-353516/00855967006011/24662116, 230701, 242161; 27-253519/00855967006028/24662719, 242171; 27-253522/00855967006141/242172; 27-253525/00855967006158/242173; 27-253528/00855967006165/242174; 27-253530/00855967006172/242175; 27-253533/00855967006189/24662733, 242176; 27-253535/00855967006196/24662735, 242177
Distribution
US Nationwide distribution in the states of VT, FL, TN, NY, GA, TX, HI.
Agency product description and boundaries

Arrow-LOK Implant, REF: 21-253519, 21-253522, 21-253525, 21-253528, 21-253530, 21-253533, 21-253535, 21-253540, 21-253545, 21-253550, 21-353513, 21-353516, 27-253519, 27-253522, 27-253525, 27-253528, 27-253530, 27-253533, 27-253535. Implant indicated for fixation of osteotomies, arthrodeses and reconstruction in the lesser toes following corrective procedures.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Implant did not undergo required gamma radiation dose audit. A potential reduction in sterility assurance could lead to patient infection.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Arrow-LOK Implant recall status

3. Open the official notice or contact

Official recall ID
Z-2958-2026
Brand
All Arrowhead DE recalls
Classification
Class II
Units affected
4,171
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Peachtree City, GA, United States
Firm address
401 Westpark Ct Ste 200, 30269-3572

Identifying details (verbatim from the notice)

Product codes / lots
REF/UDI-DI/Lot: 21-253519/00855967006035/24662119, 242162; 21-253522/00855967006042/24662122, 242163, 220702; 21-253525/00855967006059/24662125, 220703, 242164; 21-253528/00855967006066/24662128, 242165, 230704; 21-253530/00855967006073/242166, 230705, 220705; 21-253533/00855967006080/24662133, 242167, 230706; 21-253535/00855967006097/24662135, 242168, 230707; 21-253540/00855967006110/242169; 21-253545/00855967006127/242170; 21-253550/00855967006134/24662150; 21-353513/00855967006004/24662113, 242160; 21-353516/00855967006011/24662116, 230701, 242161; 27-253519/00855967006028/24662719, 242171; 27-253522/00855967006141/242172; 27-253525/00855967006158/242173; 27-253528/00855967006165/242174; 27-253530/00855967006172/242175; 27-253533/00855967006189/24662733, 242176; 27-253535/00855967006196/24662735, 242177
Distribution
US Nationwide distribution in the states of VT, FL, TN, NY, GA, TX, HI.
Firm notified the public via
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
FDA event ID
99383
Recall initiated
June 23, 2026
FDA classified
August 14, 2026
FDA report date
August 26, 2026

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 6:00 PM UTC. Record retrieved October 22, 2026 at 6:00 PM UTC.