Fresh Recalls

FDAClass I · Medical device · Recall Z-3090-2026 · Recalled September 16, 2026

Bravo CF capsule delivery device recall — Given Imaging Ltd.

Agency report date: September 16, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Bravo CF capsule delivery device
Company / brand
Given Imaging
Lot / serial / date codes
Product Number FGS-0635 UDI-DI (GTIN) 07290101369707 Product Number FGS-0636 UDI-DI (GTIN) 07290101369714 Expiration dates prior to and including October 20, 2027 (lot numbers from 64553F up to and including 68912F)
Distribution
Worldwide distribution. US nationwide including Puerto Rico; Argentina, Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Czech Republic, Dominican Republic, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Kazakhstan, Kenya, Luxembourg, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, and United Kingdom.
Agency product description and boundaries

Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Capsule may fail to attach to the esophagus or detach from the delivery device. Risks include patient aspiration/inhalation, perforation of esophagus, obstruction of the airway, hemorrhage/blood loss/bleeding, laceration of the esophagus, a delay in diagnosis, and foreign bodies remaining in the patient.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class I: most serious — reasonable probability of serious health consequences or death.

Affected product: Bravo CF capsule delivery device recall status

3. Open the official notice or contact

Official recall ID
Z-3090-2026
Brand
All Given Imaging recalls
Classification
Class I
Units affected
127,416
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Yokne'Am Ilit, Israel
Firm address
New Industrial Park, 2, Hacarmel St.; Ta Do'Ar 258, N/A

Identifying details (verbatim from the notice)

Product codes / lots
Product Number FGS-0635 UDI-DI (GTIN) 07290101369707 Product Number FGS-0636 UDI-DI (GTIN) 07290101369714 Expiration dates prior to and including October 20, 2027 (lot numbers from 64553F up to and including 68912F)
Distribution
Worldwide distribution. US nationwide including Puerto Rico; Argentina, Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Czech Republic, Dominican Republic, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Kazakhstan, Kenya, Luxembourg, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, and United Kingdom.
Firm notified the public via
Letter
FDA event ID
99626
Recall initiated
August 6, 2026
FDA classified
September 10, 2026
FDA report date
September 16, 2026

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 6:00 PM UTC. Record retrieved October 22, 2026 at 6:00 PM UTC.

Report a data error on this page