Fresh Recalls

FDAClass I · Medical device · Recall Z-3139-2026 · Recalled September 30, 2026

INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP recall — CooperSurgical, Inc.

Agency report date: September 30, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP
Company / brand
CooperSurgical
Lot / serial / date codes
1. Model Number: 44-2707. UDI-DI: (01)20888937009433. Lot Code: 550019722. 2. Model Number: 44-2709. UDI-DI: (01)60888937009448. Lot Codes: 550031617, 550033781, 550031621. 3. Model Number: 44-2710. UDI-DI: (01)20888937009457. Lot Codes: 550031615, 550031616, 550031620, 550034760. 4. Model Number: 44-2712. UDI-DI: (01)20888937009464. Lot Codes: 550031611, 550031614, 550033779, 550031619. 5. Model Number: 44-2715. UDI-DI: (01)20888937009471. Lot Codes: 550031618, 550033780, 550034763.
Distribution
Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
Agency product description and boundaries

INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class I: most serious — reasonable probability of serious health consequences or death.

Affected product: INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP recall status

3. Open the official notice or contact

Official recall ID
Z-3139-2026
Brand
All CooperSurgical recalls
Classification
Class I
Units affected
2,643 boxes (13,215 units)
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Trumbull, CT, United States
Firm address
95 Corporate Dr, 06611-1350

Identifying details (verbatim from the notice)

Product codes / lots
1. Model Number: 44-2707. UDI-DI: (01)20888937009433. Lot Code: 550019722. 2. Model Number: 44-2709. UDI-DI: (01)60888937009448. Lot Codes: 550031617, 550033781, 550031621. 3. Model Number: 44-2710. UDI-DI: (01)20888937009457. Lot Codes: 550031615, 550031616, 550031620, 550034760. 4. Model Number: 44-2712. UDI-DI: (01)20888937009464. Lot Codes: 550031611, 550031614, 550033779, 550031619. 5. Model Number: 44-2715. UDI-DI: (01)20888937009471. Lot Codes: 550031618, 550033780, 550034763.
Distribution
Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
Firm notified the public via
Letter
FDA event ID
99643
Recall initiated
August 11, 2026
FDA classified
September 18, 2026
FDA report date
September 30, 2026

More CooperSurgical recalls

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 6:00 PM UTC. Record retrieved October 22, 2026 at 6:00 PM UTC.

Report a data error on this page