Beckman Coulter Mishima K.K recalls
Recall found. Beckman Coulter Mishima K.K is named in FDA Class II recall Z-3332-2026 covering Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 RE.
Hazard: Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results. Identifying details: Products containing affected software: Item Numbers / UDI-DI codes: Item Number: B96692 UDI-DI code: 14987666542965 Item Number: B96693 UDI-DI code: 14987666542972 Item Number: B12183 UDI-DI code: 14987666540671 Item Number: N3151400 UDI-DI code: Due to the age of the device and the fact that it has exceeded its storage period, the UDI number is not available.). Status: Ongoing.
Models with recalls
All Beckman Coulter Mishima K.K recalls
FDA Class II · Medical device · October 7, 2026
Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the …
Official ID Z-3332-2026 ·
Official record ↗
Compiled from official public-domain government data. Last checked October 22, 2026 at 12:00 PM UTC.