Fresh Recalls

FDAClass II · Medical device · Recall Z-3332-2026 · Recalled October 7, 2026

Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 RE recall — Beckman Coulter Mishima K.K.

Agency report date: October 7, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 RE
Company / brand
Beckman Coulter Mishima K.K
Lot / serial / date codes
Products containing affected software: Item Numbers / UDI-DI codes: Item Number: B96692 UDI-DI code: 14987666542965 Item Number: B96693 UDI-DI code: 14987666542972 Item Number: B12183 UDI-DI code: 14987666540671 Item Number: N3151400 UDI-DI code: Due to the age of the device and the fact that it has exceeded its storage period, the UDI number is not available.)
Distribution
Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.
Agency product description and boundaries

Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only.

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 RE recall status

3. Open the official notice or contact

Official recall ID
Z-3332-2026
Brand
All Beckman Coulter Mishima K.K recalls
Classification
Class II
Units affected
492 units
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
Sunto-Gun, Japan
Firm address
454-32, Higashino; Nagaizumi-Cho, N/A

Identifying details (verbatim from the notice)

Product codes / lots
Products containing affected software: Item Numbers / UDI-DI codes: Item Number: B96692 UDI-DI code: 14987666542965 Item Number: B96693 UDI-DI code: 14987666542972 Item Number: B12183 UDI-DI code: 14987666540671 Item Number: N3151400 UDI-DI code: Due to the age of the device and the fact that it has exceeded its storage period, the UDI number is not available.)
Distribution
Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.
Firm notified the public via
Letter
FDA event ID
99588
Recall initiated
August 3, 2026
FDA classified
September 30, 2026
FDA report date
October 7, 2026

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 12:00 PM UTC. Record retrieved October 22, 2026 at 12:00 PM UTC.