FDA Class II · Medical device · July 22, 2026

Photera 600 3-D recall — Daavlin Distributing Company

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

Official ID Z-2744-2026 · Official record ↗

FDA Class II · Medical device · July 22, 2026

Photera 600 recall — Daavlin Distributing Company

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

Official ID Z-2743-2026 · Official record ↗

FDA Class II · Medical device · July 22, 2026

Photera 400 recall — Daavlin Distributing Company

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

Official ID Z-2742-2026 · Official record ↗

FDA Class II · Medical device · June 10, 2026

Phothera 200 recall — Daavlin Distributing Company

Some of the Calibration/Output Certificates had the outputs reversed.

Official ID Z-2295-2026 · Official record ↗