FDA Class II · Medical device · July 22, 2026
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Official ID Z-2744-2026 ·
Official record ↗
FDA Class II · Medical device · July 22, 2026
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Official ID Z-2743-2026 ·
Official record ↗
FDA Class II · Medical device · July 22, 2026
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Official ID Z-2742-2026 ·
Official record ↗
FDA Class II · Medical device · June 10, 2026
Some of the Calibration/Output Certificates had the outputs reversed.
Official ID Z-2295-2026 ·
Official record ↗