Fresh Recalls

Photera 600: recall status

Notice names this product. Daavlin Distributing Company’s Photera 600 is named in FDA Class II recall Z-2743-2026. Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies. Agency-reported hazard: Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000). Identifying details: Serial Numbers: 6-09033NB4CL, 6-09037NB4CL, 6-09031NB4CL, 6-09538NB4CL, 6-09038NB4CL, 6-09039NB4CL, 6-09040NB4CL, 6-09041NB4CL, 6-09035NB4CL, 6-09034NB4CL, 6-09030NB4CL, 6-09026NB4CL, 6-09023NB4CL, 6-09028NB4CL, 6-09025NB4CL, 6-09043NB4CL, 6-09029NB4CL, 6-09044NB4CL, 6-09022NB4CL, 6-09027NB4CL, 6-09036NB4CL, 6-09045NB4CL, 6-09024NB4CL, 6-07827NB4CL, 6-09539NB4CL, 6-09042NB4CL.. Units affected: 26 units. Status: Ongoing.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Photera 600
Company / brand
Daavlin Distributing Company
Lot / serial / date codes
Serial Numbers: 6-09033NB4CL, 6-09037NB4CL, 6-09031NB4CL, 6-09538NB4CL, 6-09038NB4CL, 6-09039NB4CL, 6-09040NB4CL, 6-09041NB4CL, 6-09035NB4CL, 6-09034NB4CL, 6-09030NB4CL, 6-09026NB4CL, 6-09023NB4CL, 6-09028NB4CL, 6-09025NB4CL, 6-09043NB4CL, 6-09029NB4CL, 6-09044NB4CL, 6-09022NB4CL, 6-09027NB4CL, 6-09036NB4CL, 6-09045NB4CL, 6-09024NB4CL, 6-07827NB4CL, 6-09539NB4CL, 6-09042NB4CL.
Distribution
US Nationwide distribution.
Agency product description and boundaries

Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

Brand
Daavlin Distributing Company

Recalls naming this model

FDA Class II · Medical device · July 22, 2026

Photera 600 recall — Daavlin Distributing Company

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

Company / brand: Daavlin Distributing Company recalls

Affected product: Photera 600 recall status

Official ID Z-2743-2026 · Official record ↗

See all Daavlin Distributing Company recalls →

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from connected official public-domain government data. Sources last checked October 22, 2026 at 6:00 PM UTC.