Dexcom recalls
4 recent recalls on record for Dexcom in CPSC, NHTSA, and FDA data as of July 24, 2026 at 1:09 AM UTC.
Models with recalls
All Dexcom recalls
FDA Class II · Medical device · July 1, 2026
Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies …
Official ID Z-2574-2026 ·
Official record ↗
FDA Class II · Medical device · May 20, 2026
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app…
Official ID Z-2165-2026 ·
Official record ↗
FDA Class II · Medical device · May 20, 2026
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app…
Official ID Z-2164-2026 ·
Official record ↗
FDA Class II · Medical device · May 20, 2026
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app…
Official ID Z-2163-2026 ·
Official record ↗
Compiled from official public-domain government data. Last checked July 24, 2026 at 1:09 AM UTC.