Fresh Recalls

FDAClass II · Medical device · Recall Z-2574-2026 · Recalled July 1, 2026

Brand Name recall — Dexcom, Inc.

Agency report date: July 1, 2026.

1. Compare your item

This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.

Product named
Brand Name
Company / brand
Dexcom
Lot / serial / date codes
Lot Code: UDI-DI: 00386270004109 Lot Numbers: 1725204004 and 1725069002
Distribution
Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM App user performance data. Based on App user performance data, Lot 1725204004 has sensor session starts in the United States. Based on App user performance data, Lot 1725069002 has sensor session starts in the United States and the countries of Australia, Austria, Canada, Czech Republic, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, South Korea, Kuwait, Netherlands, Poland, Portugal, Qatar, Singapore, Spain, Switzerland, Turkey, United Arab Emirates,and the United Kingdom.
Agency product description and boundaries

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System

Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.

If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.

2. Agency hazard and consumer action

FDA reports: Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.

No consumer remedy instructions are supplied in this record. Check the official notice and listed company contact for the agency’s current instructions.

Class II: may cause temporary or medically reversible health consequences.

Affected product: Brand Name recall status

3. Open the official notice or contact

Official recall ID
Z-2574-2026
Brand
All Dexcom recalls
Classification
Class II
Units affected
N/A
Recall status
Ongoing
Initiated
Voluntary: Firm initiated

Affected products

Companies & origin

Recalling firm location
San Diego, CA, United States
Firm address
6340 Sequence Dr, N/A, 92121-4356

Identifying details (verbatim from the notice)

Product codes / lots
Lot Code: UDI-DI: 00386270004109 Lot Numbers: 1725204004 and 1725069002
Distribution
Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM App user performance data. Based on App user performance data, Lot 1725204004 has sensor session starts in the United States. Based on App user performance data, Lot 1725069002 has sensor session starts in the United States and the countries of Australia, Austria, Canada, Czech Republic, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, South Korea, Kuwait, Netherlands, Poland, Portugal, Qatar, Singapore, Spain, Switzerland, Turkey, United Arab Emirates,and the United Kingdom.
Firm notified the public via
Letter
FDA event ID
98968
Recall initiated
May 27, 2026
FDA classified
June 24, 2026
FDA report date
July 1, 2026

Other medical device recalls from around the same time

Verify with the official sources:

Found something wrong or missing? Report a data error.

Source: U.S. Food & Drug Administration. Agency excerpts come from the public data feed and may omit tables or later changes; the linked government source is authoritative. Source last checked October 22, 2026 at 6:00 PM UTC. Record retrieved September 29, 2026 at 6:00 PM UTC.