Fresh Recalls

⚠ Data freshness notice: the most recent refresh of NHTSA was incomplete, so a very recent recall may be missing. The stored records may also have changed at the agency since the last successful check; for anything time-critical, confirm with the official sources linked below.

Greiner Bio-One North America recalls

Recall found. Greiner Bio-One North America is named in FDA Class II recall Z-0200-2027 covering Brand Name. Hazard: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia. Identifying details: Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12. Status: Ongoing.

Models with recalls

Brand Name (1)

All Greiner Bio-One North America recalls

FDA Class II · Medical device · October 28, 2026

Brand Name recall — Greiner Bio-One North America, Inc.

Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased …

Official ID Z-0200-2027 · Official record ↗

Verify with the official sources:

Found something wrong or missing? Report a data error.