⚠ Data freshness notice: the most recent refresh of NHTSA was incomplete, so a
very recent recall may be missing. The stored records may also have changed at the agency since the last successful check;
for anything time-critical, confirm with the official sources linked below.
Greiner Bio-One North America recalls
Recall found. Greiner Bio-One North America is named in FDA Class II recall Z-0200-2027 covering Brand Name.
Hazard: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia. Identifying details: Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12. Status: Ongoing.
Models with recalls
All Greiner Bio-One North America recalls
FDA Class II · Medical device · October 28, 2026
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased …
Official ID Z-0200-2027 ·
Official record ↗
Compiled from official public-domain government data. Last checked November 25, 2026 at 12:00 AM UTC.