⚠ Data freshness notice: the most recent refresh of NHTSA was incomplete, so a
very recent recall may be missing. The stored records may also have changed at the agency since the last successful check;
for anything time-critical, confirm with the official sources linked below.
Brand Name: recall status
Notice names this product. Greiner Bio-One North America’s Brand Name is named in FDA Class II recall Z-0200-2027.
Your particular item is not yet verified. Compare the identifiers and exclusions in the notice before deciding whether it applies.Agency-reported hazard: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia. Identifying details: Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12. Units affected: 860,450. Status: Ongoing.
1. Compare your item
This notice names a product; your particular item is not yet verified. Compare every applicable model, serial, lot, production-date or model-year boundary below.
Product named
Brand Name
Company / brand
Greiner Bio-One North America
Lot / serial / date codes
Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12
Distribution
U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
Agency product description and boundaries
Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)
Still to verify: match the label on your item with the full official notice, including any exclusions, tables and photos. Some identifiers appear only on the agency website and are absent from this data feed. A missing field does not mean all versions are affected.
If your lot or date falls outside a stated range, this notice has not established that your item is affected. This is not a finding about other recalls or product safety.
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased …