Fresh Recalls

Laborie Medical Technologies recalls

Recall found. Laborie Medical Technologies is named in FDA Class II recall Z-2669-2026 covering Solar GI System (Water Perfusion). Hazard: Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed. Identifying details: 1. Catalog Number: G3-6; UDI-DI: Primary UDI-DI: 08717775955047; Basic UDI-DI: 87177759500G3G8; All serial numbers; 2. Catalog Number: G3-7; UDI-DI: Primary UDI-DI: 08717775955054; Basic UDI-DI: 87177759500G3G8; All serial numbers; 3. Catalog Number: G3-8; UDI-DI: Primary UDI-DI: 08717775955061; Basic UDI-DI: 87177759500G3G8; All serial numbers; 4. Catalog Number: G3-12; UDI-DI: Primary UDI-DI: 08717775955078; Basic UDI-DI: 87177759500G3G8; All serial numbers; 5. Catalog Number: G3-14; UDI-DI: Primary UDI-DI: 08717775955085; Basic UDI-DI: 87177759500G3G8; All serial numbers;. Status: Ongoing.

Models with recalls

Solar GI System (Water Perfusion) (1)

All Laborie Medical Technologies recalls

FDA Class II · Medical device · July 8, 2026

Solar GI System (Water Perfusion) recall — Laborie Medical Technologies Corp

Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the h…

Official ID Z-2669-2026 · Official record ↗

Verify with the official sources:

Found something wrong or missing? Report a data error.

Compiled from official public-domain government data. Last checked October 22, 2026 at 6:00 PM UTC.